1.0 Purpose
The purpose of this procedure is to describe the setup and operation of the digital cap torque
tester in use at Ion Labs, Inc.
2.0 Scope
This procedure applies to the digital cap torque tester in use at Ion Labs, Inc.
3.0 Responsibility
4.0 It is the responsibility of QC Inspectors and Production (Packaging) personnel to follow this
procedure for all products that require cap torque testing.
5.0 Definitions
5.1 QC — Quality Control
5.2 RT — Real Time
5.3. PCW - Peak Clockwise
5.4 PCCW -— Peak Counter — Clockwise
5.5 IbFin — Pound-inch
6.0 References
6.1 Series TT01 Digital Cap Torque Tester User’s Guide
[SOP
TT S O t 1 a n D d i a g r i d ta O l p C er a a p t i T n o g r P q r u o e c e T d e u s r t e e r S B O - P 60 N 1 o Re 2 v | Page2ofs
7.0 Specifications
Hal Accuracy:
7.1.1 +0.3% of full scale
7.2 Sampling Rate:
7.2.1. 7000-Hz
7.3. Power:
7.3.1 AC or rechargeable battery
7.3.1.1 Multi-step low battery indicator is displayed, tester shuts off
automatically when power is too low.
7.4 Battery Life:
7.4.1 Backlight on: up to 7 hours of continuous use
7.4.2 Backlight off: up to 24 hours of continuous use
7.5 Measurement Units:
7.5.1 ozFin
7.5.2 IbFin
7.5.3. kgFem
7.5.4 Nm
7.5.5 Nem
7.6 Outputs:
7.6.1 USB/RS-232: Configurable up to 115,200 baud.
7.6.2 Mitutoyo: Serial BCD suitable for all Mitutoyo SPC-compatible devices.
7.6.3 Analog: +1 VCD, +0.25% of full scale at capacity.
7.6.4 General Purpose: Three open drain outputs, one input.
7.6.5 Set Points: Three open drain lines.
[SOP
Standard Operating Procedure SOP No Rev
B-601 ) Page 3 of 5
TTO1 Digital Cap Torque Tester
7.7. Configurable Settings:
dad Digital Filters
bi Pr Outputs
LAS Automatic output / data storage / zero
7.7.4 Automatic shutoff
LAD Default settings
7.7.6 Key tones
tld Backlight
7.7.8 Audio alarms
7.7.9 Passwords
7.7.10 Calibration
7.8 Safe Overload:
7.8.1 150% of full scale (display shows “OVER” at 110% and above)
7.9 Weight:
7.9.1 8.4 Ib [3.8 kg]
8.0 Controls
Primary Label Primary Function Secondary Label Secondary Function
, ; Powers the tester on and off. Press
briefly to power on, press and hold ENTER Various uses, as described in the
to power off. Active only when the following sections.
a home screen is displayed.
ZERO Zeroes the primary reading and Ma. Navigates up through the menu
peaks. (UP) and sub-menus.
; Reverts one step backwards
MENU Accesses the main menu. ESCAPE through the menu hierarchy.
MODE Toggles between measurement ~ 4 Navigates down through the menu
modes. (DOWN) and sub-menus.
Stores a value to memory and/or Toggles between clockwise and
DATA transmits the current reading to an DIRECTION counter-clockwise directions
external device, depending on while configuring set points and
configuration. other menu functions.
[SOP
Standard Operating Procedure SOP No Rev
TTO1 Digital Cap Torque Tester B-601 2 | Page 4 ofS
8.0 Procedure
8.1 To turn the power on, press ENTER.
8.2 Press the MODE until the measurement mode PCCW appears.
8.3. To the right of PCCW, the unit measurement should indicate IbFin.
Display Indicators Clockwise/counter-clockwise indicator
Set points —— ~ Peak readings (CVWW/CCW or Ist/2nd peaks)
Current or peak reading
Set point (pass/fail) indicators — Weasurement mode
Battery indicator ~ Unit of measurement
Memory storage for 1000 readings Analog load bar w/set point markers
(auto-save also possible}
Zero the reading to begin a new test
Access the testers settings menu via
an easy-to-use interface {auto-zero also possible)
Out-gush data output to a PC, printer, or Toggle between real time, peak, and 1st
other device (auto-output also possible) ets on peak/2nd peak display modes
8.4 Toconfigure set points, select Set Points from the menu.
8.4.1 Press MENU
8.4.2 Press MODE; use TM to go to Set Points
8.4.3. Two limits, high and low, are programmed in the tester, and the primary reading
is compared to these limits.
8.5 Place a bottle on the sample grippers of the torque tester and tighten the grippers using
the knob on the right of the tester. Do not over tighten.
8.6 Press ZERO on the keypad and ensure that the screen is indicating 0.00.
8.7. Grip the sample firmly by hand and gradually exert counterclockwise torque on the
bottle cap.
8.8 The torque “pound inches” reading will appear on the screen. Check the current reading
with the upper and lower pound inch limits in the upper left corner of the screen.
8.9 To turn the power off, press ENTER. When the tester is not in use, ensure that the
power is turned off.
[SOP
Standard Operating Procedure SOP No Rev ee
TT01 Digital Cap Torque Tester Bal 2 >
9.0 Revision History
| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 09/30/14 | New I 12/19/19 Scheduled review: no changes needed 19-0953 J. Kalem | 14-0774 | M. Wienke |
| 2 | 05/05/22 | Changed SOP number and logo. Updated SOP format. CC- | 22-0222 | K. Burris |