Searchable text extracted from PDF

1.0 Purpose 
 
 1.1 The objective of the Sales-Laboratory Sample Submission process is to ensure
 requirements of customers, regulatory authorities, and accreditation organizations are
 
 met. Another objective is to ensure information provided to customers regarding Ion
 Labs’ capability and testing requirements are accurate and communicated clearly.
 
 1.2 The purpose of this procedure is to describe the process requirements including
 responsibilities and controls for ensuring all test activities are completed per customer
 requests. This procedure describes the requirements for communicating the scope of
 
 accreditation to potential customers. This procedure also describes the process for
 collecting, reviewing, and responding to customer feedback. 
 
 2.0 Scope 
 
 2.1 This procedure is applicable to all ISO 17025 accredited test activities performed by Ion
 
 Labs. 
 
 3.0 Responsibility 
 
 Suk 
 Management is responsible for ensuring tests are performed as requested by customers.
 Management includes the Lab Director and Analytical Development Manager.
 
3.2 Management is responsible for reacting and responding to customer feedback regarding

 lab activities including complaints and notifications of suspected nonconforming test
 activities. Management is also responsible for determining the customers’ level of
 satisfaction and their perception of Ion Lab’s laboratory performance. When complaints
 are received, Management will be responsible for all decisions at all levels of the handling
 process for complaints in accordance with the QS-112 Core Quality Systems and Quality
 Events. 
 
 3.3 All personnel participating is Sales-Laboratory Sample Submission activities are
 responsible for ensuring that information communicated to customers or potential
 customers, correctly identifies the test services accredited to ISO 17025.
 
3.4 Sales personnel are responsible for communicating with customers to determine their

 requirements and ensuring they are documented on Ion Lab’s quotes and Sales Orders.
 

[SOP 

 Standard Operating Procedure SOP No | Rev No | _ Page 
 Sales - Laboratory Sample Submission D-119 0 2 of 10 
 
 The completed quote documents Ion Lab’s understanding of customer’s requirements.
 
 3.3 The Sales Manager and Lab Director are responsible for ensuring that this procedure is
 accurate, understood and implemented effectively. No changes may be made to this
 procedure without the authorization of the Sales Manager and Lab Director.
 
 4.0 Definition 
 
 4.1 Customer — A person or organization that could or does receive a product or a service
 that is intended for or required by this person or organization. A customer can be internal
 or external to the organization. 
 
 4.1.1 Examples include but are not limited to customer, client, end-user, retailer,
 
 receiver of product or service from an internal process beneficiary and purchaser.
 
 5.0 References 
 
 5.1 D-201, SOP, QC Laboratory Sample Log Book Recording 
 
 5.2 D-201-F1, Form, QC Laboratory Sample Log 
 
 5.3 C-502, SOP, Record Storage, Retention, and Destruction 
 
 5.4 QS-108, SOP, Corrective and Preventive Action 
 
 5.5 E-601, SOP, Vendor Qualification 
 
 5.6 D-117, SOP, Laboratory Purchasing and Receiving 
 
 5.7 D-103, SOP, Analytical Method Validation 
 
 6.0 Procedure 
 
 6.1 Inputs to the Sales-Laboratory Sample Submission process include, but are not limited
 to: 
 
 ® Customer requests for test services, 
 
 6 Test deviation requests, 
 
 @ Accreditation policies, 
 
 @ Customer feedback, including complaints. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | RevNo |_ Page 
 Sales - Laboratory Sample Submission D-119 0 3 of 10 
 
 6.2 General 
 
 6.2.1 Customers may be internal customers of Ion Labs and the same process applies.
 
 6.2.2 If Ion Labs subcontracts testing, whether planned or due to unforeseen
 circumstances, this work is awarded to a competent subcontractor that is
 accredited to ISO 17025 for the required test. Customers are advised in writing
 
 of the proposed use of subcontractors, and written approval of the customers is
 required before proceeding with the tests. The written communications occur via
 email. 
 
 6.2.2.1 Ion Labs remains responsible to the customer for any subcontracted
 work, except in cases where the customer or regulatory authorities
 
 specify the use of subcontractors. 
 
 6.2.2.2. A list of approved subcontractors and the subcontracted work is
 maintained in accordance with the E-601 Vendor Qualification and D-
 117 Laboratory Purchasing and Receiving procedure. 
 
 6.2.3 Ion Labs utilizes laboratory-developed methods which modify established test
 methods to accommodate the purpose and intent of the published established test
 methods. Deviations to established test methods will be identified and
 documented as described in section 6.3.10 below. Test methods selected will be
 established methods published either in international, regional, or national
 standards, or by reputable technical organizations, in relevant scientific texts or
 
 journals, or by the original equipment manufacturer. Lab-developed methods will
 be based on published methods in accordance with the D-103 Analytical Method
 Validation. 
 
 6.2.4 Confidentiality 
 
 6.2.4.1 Ion Lab’s legally enforceable responsibilities for the management of
 
 information obtained or created during the performance of test activities
 is documented within a Customers Purchase Order executed with each
 customer. 
 
 6.2.4.2 Ion Labs does not typically place information obtained or created during
 the performance of test activities into the public domain. In the event
 
 Ion Labs intends to place information into the public domain, affected
 customers will be notified in advance in writing of this intention,
 including details regarding the information to be disclosed, and the
 methods of disclosure. Information placed in the public domain will not
 violate the legally enforceable responsibilities documented in
 
 
 

[SOP 

 Standard Operating Procedure SOP No | RevNo | _ Page 
 Sales - Laboratory Sample Submission D-119 0 4 of 10 
 
 Customers Purchase Order established with affected customers.
 
 6.2.4.3 As documented on the Customers Purchase Order and the Sales Order,
 all information (including materials and results) will be considered
 proprietary information and will be regarded as confidential between the
 customer and Ion Labs. Where agreed by the customer and Ion Labs,
 
 customers may place information generated during test activities into
 the public domain within the limits established in the Purchase Order.
 
 6.2.4.4 In the event Ion Labs is compelled to release information by court order,
 information will be provided as required. Unless prohibited by law or
 court order, the customer or individual concerned will be notified of the
 
 information provided in association with the release of information.
 
 6.2.4.5 In the event external sources provide information regarding customers
 (for example, in the event of complaints or investigations overseen by
 regulators), this information will remain confidential between the
 
 customer and Ion Labs. The source of this information will be
 confidential to Ion Labs, and will not be disclosed to the customer,
 unless agreed by the source in writing. 
 
 6.2.4.6 Personnel acting on Ion Labs’ behalf (including contractors, and other
 personnel of external bodies) will keep confidential all information
 obtained or created during the performance of laboratory activities,
 
 except where required by law to disclose such information.
 Communication of these requirements to affected parties will occur as
 described in E-601 Vendor Qualification and D-117 Laboratory
 Purchasing and Receiving procedure. 
 
 6.2.5 Decision Rule 
 
 6.2.5.1 The decision rule is the rule that describes how measurement uncertainty
 is accounted for when stating conformity with a specified requirement.
 
 6.2.5.2 As stated on the Customer’s Purchase Order, the decision rule for the
 test’s conformity or nonconformity does not consider the measurement
 uncertainty for the test activities performed. The decision is based on
 
 the actual results. The statement on the form serves as the
 communication to the customer. 
 
625.3 However, if the customer requests another specification or standard, the

 decision rule will be selected and agreed to by the customer. The agreed
 upon decision rule will be noted on a revised or a new Customer
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev No | Page 
 Sales - Laboratory Sample Submission D-119 0 5 of 10 
 
 Purchase Order. 
 
 6.2.6 Accreditation References 
 
 6.2.6.1 Management will ensure the accreditation body’s policy concerning
 their requirements when making reference to the accredited status will
 be strictly adhered to. ANAB Accreditation Requirement AR2201,
 
 “Control and Use of Accreditation Symbol provides detailed
 explanations of the requirements generalized below (see the policy
 located in the Documents of External Origin folder for all requirements).
 In general, as part of ensuring compliance to the policy, Management
 
 will ensure the following during communications with potential
 customers: 
 
 e lon Labs may refer to its ANAB accreditation on test reports, on
 its website, in its general literature and promotional materials, and
 in business solicitations, under the following provisions:
 
 lon Labs may only use the accreditation symbol or reference
 O 
 to ANAB accreditation if the accreditation status is current and
 remains in good standing. 
 The accreditation symbol or reference to ANAB accreditation
 shall not be placed on products or items that Ion Labs has
 
 tested. 
 
 Ion Labs shall not use the accreditation symbol or refer the
 accreditation in any way that may be misleading or
 misrepresent the accreditation status. 
 
 Ion Labs shall ensure claims of accreditation are related only
 to activities within their scope of accreditation and not
 
 associated with other activities in which Ion Labs may be
 involved. 
 
 [on Labs shall ensure that the use of the accreditation symbol
 or reference to the accreditation status is not affixed to a
 product (or part of it) or used in any way to imply that a product
 
 has been certified. 
 
 Ion Labs shall not use the accreditation symbol or refer to the
 accreditation in any manner that gives the impression that
 ANAB accepts responsibility for the results, or for any opinion
 
 
 

[SOP 

 Standard Operating Procedure SOP No | RevNo | _ Page 
 Sales - Laboratory Sample Submission D-119 0 6 of 10 
 
 or interpretation derived from those results, or that ANAB
 approves of those results. 
 
 o Itis lon Labs’ responsibility to use the accreditation symbol or
 any materials representing our accreditation in such a way as
 to enhance the reputation and value of accreditation for all
 
 stakeholders. 
 
 o The ANAB accreditation symbol always shall be used in its
 original, designed proportions and shall not be modified,
 distorted, compressed, or stretched in any way. It will not be
 
 sized in a way that renders it unreadable. 
 
 o lon Labs will use the accreditation symbol as provided by
 ANAB, at Ion Labs discretion. 
 
 o lon Labs shall not imply or mislead the scope of its accredited
 activities using certificates or reports, enclosures, stationery,
 or other materials. Ion Labs shall not lead any user of the
 
 results or interested party to believe that the work is covered
 by the scope of accreditation when it is not. 
 
 6.2.6.2 All records associated with this process, including accepted Customer
 Purchase Orders, communications regarding the customers’
 requirements, approved deviations, and feedback indicating customer
 
 satisfaction and responses are retained per SOP C-502 Record Storage,
 Retention, and Destruction. 
 
 6.3 Sales Order Establishment and Review 
 
 6.3.1 Laboratory activities will be performed based on documented quotes and
 purchase orders established with customers describing the tests to be performed
 
 and the services to be provided. Sales personnel may initiate quotes with
 customers. Lab personnel review and approve the quotes to ensure customer
 requirements have been adequately defined, documented, understood, and Ion
 Labs has the capability and resources to meet customer requirements. Special
 requirements beyond those documented in the quote may be documented in
 
 emails or other correspondence with the Sales personnel and/or Lab personnel
 and retained with the orders’ records. 
 
 6.3.2 Requests for quotes will be reviewed to identify risks to impartiality.
 Management may be contacted prior to processing the request to determine if
 risks to impartiality potentially exist based on the customer or the service
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev No Page 
 Sales - Laboratory Sample Submission D-119 7 of 10 
 
 requested. If no risks to impartiality are determined by Management, the request
 
 may be processed as described below, and the quote will document no risk to
 impartiality was determined. Ifa risk to impartiality exists, the Lab Director will
 take steps to address the risk and communicate with the customer as appropriate.
 Risks to impartiality may arise from Jon Labs’ activities, from Ion Labs’
 
 relationships with other organizations, or from the relationships of Ion Labs’
 personnel. If ongoing actions will be required to address the risks, these actions
 will be documented on the Risks and Opportunities worksheet of the LMS
 Planning Tool. 
 
 6.3.3 Sales personnel will ensure the quote includes information regarding appropriate
 
 test service(s) to satisfy the customer’s requirements. All tests will be performed
 using established test methods (Standard Test Procedures) relevant to the
 customers' needs and consistent with requirements specified by the customer. For
 ISO 17025 accredited tests, the quote must indicate if the customer wants the
 
 accredited test or not. 
 
 6.3.3.1. Accredited tests will provide the customer with all required information
 including the accreditation logo on the test report. Test Reports for
 unaccredited tests provide traceability to a national or international
 standard but may not disclose the measurement uncertainty and will not
 
 include the accreditation logo or any reference to the accreditation status
 of the Lab. 
 
 6.3.4 If the customer does not specify the established test method to be used, Lab
 personnel will review information regarding the customer’s intended outcome for
 the requested tests and if the Lab can satisfy the customer’s requirements, an
 
 appropriate test method will be selected within the scope of Ion Labs’ tests. The
 Lab Director or Analytical Development Manager indicates the method selected
 by providing the information to Sales personnel to document it in the quote.
 
 6.3.5 Ifat any time it is learned the method and/or laboratory activities requested by the
 customer are found to be inappropriate, out of date, or outside the capability of
 
 Ion Labs, the Lab Director or Analytical Development Manager will inform the
 customer of Ion Labs’ capabilities and describe why the methods and/or
 laboratory activities requested cannot be satisfied. This includes requests for
 quotes indicating accreditation for tests that are not in Ion Labs’ scope of
 
 accreditation. 
 6.3.6 As stated on the quote, the decision rule for the chosen test does not consider the
 
 measurement uncertainty of the test performed. The decision is based on the
 actual results. The statement on the form serves as the communication to the
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev No Page 
 Sales - Laboratory Sample Submission D-119 8 of 10 
 
 customer. However, if the customer requests another specification or standard,
 
 the decision rule will be selected and agreed to by the customer. The agreed upon
 decision rule will be recorded on the quote and Test Report. 
 
 6.3.7 The Lab Director or Sales personnel will ensure any differences between the
 customer’s requirements and Ion Labs’ capabilities are resolved before the quote
 
 is sent to the customer and is accepted or before any laboratory activities
 commence. This will ensure the test activities will be acceptable to both the
 customer and Ion Labs. 
 
 6.3.8 While reviewing, evaluating, and approving quotes with customers, Ion Labs staff
 members will cooperate with customers (and their representatives as necessary)
 
 to clarify the customer’s request and describe Ion Lab’s capabilities.
 
 6.3.9 Where the customer has specified performance monitoring of laboratory activities
 in relation to the work performed, Ion Labs will agree to the specified monitoring
 unless the monitoring activities will cause deviation from the established test
 method, pose a risk to personnel safety, or negatively affect the outcome of
 
 laboratory activities. 
 
 6.3.10 If laboratory activities deviate from the accepted test method, customer
 requirements, and/or the Sales Order, details regarding the deviation will be
 documented on the test paperwork and the Test Report. The test paperwork will
 document the deviated requirement, provide a technical justification for the
 
 deviation, include a record of authorization for the deviation from Ion Labs’
 Management, and include evidence of acceptance of the deviation by the
 customer. The Lab Director will evaluate deviations requested by the customer
 to ensure they will not impact the integrity of the laboratory or the validity of the
 
 results. The customer will be informed of the decision. 
 
 6.3.11 Approval of Customer’s Purchase Orders will be indicated by the creation of the
 Sales Order. The Quote and the Customer’s Purchase Orders will be retained as
 records of the requirements review and Ion Lab’s capability of satisfying the
 customer requirements. 
 
 6.3.12 If pertinent discussions (determined at the discretion of the Lab personnel based
 
 on impact of test results or service delivery) were held relating to the customer’s
 requirements or the results of the laboratory activities, records of these
 discussions will be retained Sales Order. 
 
 6.3.13 Ifthe customer amends their Purchase Order or notifies the Lab for an amendment
 after test activities have started, the process described above will be repeated to
 ensure the change can be accommodated and all affected personnel have been
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev No | _ Page 
 Sales - Laboratory Sample Submission D-119 0 9 of 10 
 
 notified. 
 
 6.4 Customer Feedback and Complaints 
 
 6.4.1 Management will seek feedback from customers, both positive and negative,
 through information gathered during direct personal contact with customers
 (including but not limited to in-person meetings, emails, and participation in
 
 industry events). Customer feedback will be recorded on the Customer Feedback
 Log as evidence of analyzing and using feedback to improve the management
 system, laboratory activities, and customer service. 
 
 6.4.2 When the feedback received is determined to be a lab or testing complaint
 regarding the performance or outcomes of laboratory activities, the QS-112 Core
 
 Quality Systems and Quality Events process will be used. 
 
 6.4.3 Customer feedback which will be analyzed and used for improvement will be
 based upon performance factors including (but not limited to) quality, timeliness
 of reported results, and accuracy and clarity of results. 
 
 6.4.4 Information regarding customer feedback will be submitted to Management
 
 during management review (see the Management procedure). Management will
 also review customer feedback as it becomes available and will react
 appropriately according to the information received, initiating corrective actions
 as appropriate as described in the QS-108 Corrective and Preventive Action.
 
 6.5 Record Management 
 
 6.5.1 Records are maintained per SOP C-502 Record Storage, Retention, and
 
 Destruction. 
 
 6.6 Process Outputs 
 
 6.6.1 Outputs to the Sales -Laboratory Sample Submission process include, but are not
 limited to: 
 
 ® Accepted orders for test services, 
 
 ® Approved deviations, 
 
 @ Positive and negative feedback received from customers and follow-up
 actions, including corrective actions where necessary. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | RevNo |_ Page 
 Sales - Laboratory Sample Submission D-119 0 10 of 10 
 
 8.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 10/19/21 | New N/A J. Sassman | - | - |