D-706

Use and Calibration of pH Meters

Section D — Laboratory Operations and Specifications Revision 6 8 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 This procedure gives guidelines on the calibration and use of the Bench 700 Series pH meter
 with ATC made by Oakton and the Mettler Toledo SevenCompactTM $220 Series pH meter.
 
 2.0 Scope 
 
 This procedure gives general guidance for the calibration and use of any pH meter. This
 
 procedure also gives specific instructions on calibration and use of an Oakton Bench 700 Series
 
 pH meter and the Mettler Toledo SevenCompactTM S220 Series pH meter. This procedure
 complies with USP <791> pH. 
 
 3.0 Responsibility 
 
 31 It is the responsibility of QC and Analytical Chemists to follow this procedure.
 
 3.2 It is the responsibility of QC Laboratory Management to implement this procedure and
 
 to ensure that the procedure is being followed. 
 
 3.3. It is the responsibility of QC Laboratory Management to keep this procedure aligned
 
 with current practices. 
 
 4.0 Definitions 
 
 4.1 IPA — Isopropyl alcohol 
 
 4.2 ATC-— Automatic Temperature Correction 
 
 4.3 QC — Quality Control 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use and Calibration of pH Meters D-706 6) (base ok! 
 
 5.0 References 
 
 5.1 USP <791> pH 
 
 5.2 D-706-F1, Form, pH Measurement Test Ticket 
 
 5.3. Mettler Toledo SevenCompactTM pH Ion Metter S220 Operating Instructions, 2011
 
 5.4 Quick Guide Bench 700 Series pH/mV/Conductivity/TDS/Dissolved Oxygen/°C/'F,
 
 Oakton, 5/11, Rev 0 
 
 5.5 C-501, SOP, Document Control Procedure 
 
 5.6 C-502, SOP, Record Storage, Retention, and Destruction 
 
 6.0 Equipment 
 
 6.1 Oakton Bench 700 Series pH Meter with All-in-One pH/ATC Probe, Sealed/SJ/Epoxy.
 
 6.2 Mettler Toledo SevenCompactTM $220 pH/Ion meter with Mettler Toledo InLab®
 
 Expert Pro-ISM pH probe. 
 
 7.0 Procedure 
 
 Td General instructions on use of a pH meter 
 
 7.1.1. When not in use the probe tip should be submerged in electrode storage solution
 
 formulated for the probe. 
 
 7.1.2 The Probes are sealed and do not require filling with electrolyte solution.
 
 7.1.3. Turn on the pH meter and allow a minimum of 15 minutes to equilibrate and
 
 stabilize. 
 
 7.1.4 The probe should be rinsed with Millipore water before use. The probe should
 
 not be wiped. If water remains on the tip of the probe it can be removed by
 gently dabbing the bottom of the probe with a Kimwipe. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use and Calibration of pH Meters D-706 6 | Page3 of7 
 
 7.1.5 For aseptic applications the probe should be rinsed with 70% IPA before use.
 The probe should not be wiped. If IPA remains on the tip of the probe it can be
 
 removed by gently dabbing the bottom of the probe with a Kimwipe.
 
 7.1.6 All buffer solutions and solutions to be buffered should be equilibrated to room
 
 temperature before testing. 
 
 7.1.7. Only NIST traceable reference buffers will be used for calibration. Buffers are
 
 available at pH 12.45, 10.01, 7.00, 4.01, and 1.68. 
 
 7.1.8 The meter should be calibrated daily or a change in calibration range.
 
 7.1.9 The pH meter should be calibrated with fresh buffer using a minimum of a 2
 
 point calibration curve the brackets the expected pH of the solution to analyzed.
 
 7.1.10 The observed pH can be temperature dependent especially at the higher pH
 
 range. To mitigate any pH variation with temperature, the temperature of the
 sample should be documented and unless otherwise stated, is required to be 25 +
 
 2G, 
 
 7.1.11 Ifthe temperature of that sample can not be in this range, the temperature of the
 
 sample must be measured by an appropriate thermometer and compared to the
 temperature reading on the ATC sensor on the pH system controller. The
 
 temperature of the thermometer and the probe should not differ by more than
 
 +2°C. If this type of reading is performed, it should be documented on the pH
 test ticket form. 
 
 7.1.12 The error measurement on the instrument is + 0.02. This should be taken into
 account when reporting pH values and comparing them to the specification.
 
 Depending on the specification, the measured value should be within
 
 specification even with the +0.02 value added. 
 
 7.2. Calibration and use of the Oakton 700 Bench Series pH Meter 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use and Calibration of pH Meters D-706 G7 aR On? 
 
 7.2.1 To calibrate the Bench 700 press the <Cal/Meas> button to enter into
 calibration mode noted by CAL on the LED screen. 
 
 7.2.2 Place the rinsed electrode in the first buffer. The Bench 700 will automatically
 recognize buffer. 
 
 7.2.3 When the LED screen blinks READY, press <Enter> to accept. 
 
 7.2.4 Rinse probe and repeat until all required reference buffers are complete.
 
 Ted After calibration is complete, press <Cal/Meas> button to return to pH mode.
 
 7.2.6 Begin measuring pH by placing the electrode tip into the solution to be
 
 measured. Adjust pH using good laboratory technique. The final measured pH
 is indicated by the stabilization of the pH and the appearance of Ready on the
 
 screen. 
 
 Tene For further quick instruction on other functions of the Oakton 700 pH Meter see
 the Quick Guide Bench 700 Series pH/mV/Conductivity/TDS/Dissolved
 
 Oxygen/°C/°F, Oakton, 5/11, Rev 0. 
 
 7.3. Operation of the Mettler Toledo SevenCompactTM pH/Ion S220 Meter 
 
 Vaid pH calibrations can be run with this meter for up to 5 points.
 
 Tidi2 For information on the setup of the pH probe see Mettler Toledo
 
 SevenCompactTM pH Ion Metter S220 Operating Instructions, 2011.
 
 7.3.3 Place the electrode in a calibration buffer and press <CAL> button.
 
 7.3.3.1. Cal 1 appears on the display. 
 
 7.3.4 The meter endpoints according to the preselected endpoint made after the signal
 
 has stabilized or after pressing the <READ> button. 
 
 7.3.4.1. The relevant buffer value is shown on the display. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use and Calibration of pH Meters D-706 Gis BABE Oly 
 
 7.3.5 Rinse the electrode with Millipore water. 
 
 7.3.6 Place the electrode in the next calibration buffer. 
 
 7.3.7 Press <CAL> button. 
 
 7.3.7.1. Cal 2 appears on the display. The meter endpoints according to the
 
 preselected endpoint mode after the signal has stabilized or after
 pressing <READ>. 
 
 7.3.7.2 The relevant buffer value is shown on the display. 
 
 Fa Repeat steps 7.3.5 to 7.3.7 for all calibration buffers. 
 
 Ten? Press <END> to end the calibration procedure. 
 
 7.3.9.1 Alternatively, the meter ends the calibration automatically when 5
 
 calibrations are performed. The offset value and slope are shown on
 
 the display. 
 
 7.3.10 Press <«> to scroll down to next data set. 
 
 7.3.11 Press <Save> to keep the calibration. 
 
 Ce ne Press <Exit> to reject the calibration. 
 
 7.3.13 To measure a sample after the calibration process is complete, place the probe
 tip in the sample and press <READ> to start a measurement. 
 
 7.3.14 The endpoint format blinks, indicating a measurement is in progress.
 
 715 The “automatic endpoint” format is used. The measurement stops automatically
 
 as soon as the Stability icon appears. 
 
 7.3.16 For further quick instruction on other functions of the Mettler Toledo
 
 SevenCompactTM S220 pH/Ion meter see Mettler Toledo SevenCompactTM pH
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Use and Calibration of pH Meters D-706 Page 6 of 7 
 
 Ion Metter S220 Operating Instructions, 2011. 
 
 7.4 Changing the Probe 
 
 7.4.1. The probe can be changed following these steps: 
 
 7.4.1.1 Remove currently attached probe by twisting interlock and removing
 
 the connection. If temperature attachment is connected, be sure to
 disconnect as well. 
 
 7.4.1.2 Attach new probe by connecting the new probe to the back on the
 meter and twisting interlock to the locked positions. If probe has a
 
 temperature auxiliary, be sure to attach as well. 
 
 7.4.1.3 The meter should recognize the name of the pH probe in the left hand
 
 corner of the display screen. 
 
 7.4.1.4 Probe should be calibrated each time it is changed. 
 
 7.5 pH measurements of gummies and other chewable gels 
 
 7.5.1. Be sure to attach appropriate probe for the measurement of gels (consult 7.4 for
 
 directions on how to change probe) 
 
 7.5.2 Carefully, insert probe into gel, ensuring at least 3/4s of the lower half of the
 probe is covered. 
 
 7.5.3 Acquire data as normal. 
 
 7.6 Calibration Notebook and Form D-706-F1 pH Measurement Test Ticket 
 
 7.6.1 This notebook will record the following information each time the pH meter is
 calibrated: 
 
 7.6.1.1 Analyst calibrating the pH meter 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Use and Calibration of pH Meters D-706 6 Page 7 of 7 
 
 7.6.1.2 Date 
 
 7.6.1.3 pH and lot number of each buffer used 
 
 7.6.2 Record results on form D-706-F1 pH Measurement Test Ticket. 
 
 77 Documentation Requirements 
 
 TAA All completed test tickets will be maintained as outlined in SOP C-501
 Document Control and SOP C-502 Record Storage, Retention, and Destruction.
 
 8.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 05/06/10 | New - - Reformatted SOP. Removed calibration buffer preparation. Gave | - | - |
| 1 | 02/18/13 | more specific instruction on setup and calibration of meter. - - Expanded information needed in calibration notebook. | - | - |
| 2 | 03/07/13 | Updated information to new pH meter. Added form. | 13-136 | B. Johns |
| 3 | 02/23/15 | Updated SOP. Added specific information on the Mettler Toledo SevenCompactTM pH meter. Biennial Review. 4 Lt204? U So p l d i a d t P e r d o S be O P p H to m r e e t f e le r c t f o c r u g rr u e m nt m p y r a u c s t e i ces and procedures. Added ilo J. Mai : gnan | 15-0171 | B. Johns |
| 5 | 04/15/21 | Updated SOP to require temperature of sample to be 25 + 2°C CC- Added documentation requirements. Added additional SOP | 21-0147 | J. Maignan |
| 6 | 10/09/23 | references. Revised test ticket to include logbook number. Updated CC- format. Updated responsibilities section. | 23-0502 | K. Burris |

 
 

[SOP 

 ION pH Measurement Test Ticket 
 
 NLuatbrsi tional Form: D-706-F1 CCR No. CC-23-0502 Revision: 7 
 
 Logbook Number: Logbook Page: 
 
 Date: 
 
 Sample Name: 
 
 Sample Identification #: 
 
 pH Meter Used: [] SevenCompactTM [] Oakton 700 
 
 pH Ion ID #: Temp probe Ion ID #: 
 
 Electrode Serial #: Cal Due Date: 
 
 Calibrated: C] Yes CL] No 
 
 $tandards Used 
 
 LI 1.68 C1 4.01 CL) 7.00 CL] 10.01 CL] 12.45 
 
 Measured pH of Sample: Tempetr a Of f ean] 
 
 Sample Specification 
 pH of sample -0.02: 
 Min: 
 Sample Specification 
 pH of sample +0.02: 
 Max: 
 Determination (circle one): Pass Fail 
 Comments: 
 
 Performed By/Date: 
 
 Reviewed By/Date: