D-725

Loss on Drying

Section D — Laboratory Operations and Specifications Revision 6 8 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 The purpose of this procedure is to define the process used for the determination of volatile
 chemical loss during drying under a specified set of conditions. 
 
 2.0 Scope 
 
 This procedure applies to the determination of loss on drying as a percentage of weight for
 powder samples. This procedure follows Loss on Drying USP<731> guidelines. Unless
 
 otherwise noted in a monograph or pharmacopeia, LOD should only be performed on solids
 
 whose melting point is above the temperature used for drying. Other forms of moisture
 determination should be used when this is not the case. 
 
 3.0 Responsibility 
 
 3.1 It is the responsibility of QC and Analytical Chemists to follow this procedure.
 
 3.2 It is the responsibility of QC Laboratory Management to implement this procedure and
 to ensure that the procedure is being followed. 
 
 3.3. «It is the responsibility of QC Laboratory and/or Analytical Development Management
 
 to keep this procedure current with latest Ion Labs practices. 
 
 4.0 Definitions 
 
 4.1 LOD — Loss on Drying 
 
 4.2 QC - Quality Control 
 
 
 

[SOP 

 Standard Operating Procedure SOPNo | _ Rev 
 D-725 6 Page 2 of 6 
 Loss on Drying 
 
 5.0 References 
 
 5.1 D-725-F1, Form, LOD Test Record 
 
 5.2. D-725-F2, Form, Halogen Moisture Analyzer Test Record 
 
 5.1 | USP<731>, Method, Loss on Drying 
 
 6.0 Procedure 
 
 6.1 Equipment 
 
 6.1.1 WWRSymphonyTM Vacuum oven or equivalent 
 
 6.1.2 Mettler Toledo Halogen Moisture Analyzer HE73 or equivalent 
 
 6.1.3 ’%4HP vacuum pump 
 
 6.1.4 Analytical Balance 
 
 6.1.5 Temperature probe 
 
 6.2 Supplies 
 
 6.2.1 30mL 34/12 Kimax Weigh Bottle or equivalent 
 
 6.2.2 Watch Glass, 12.5cm diameter 
 
 6.2.3 Mettler Toledo SmartCal Sachets 
 
 6.2.4 Aluminum Sample Pans 
 
 6.3 Testing — Drying Oven 
 
 Note: If a monograph is available to give conditions for testing LOD for a material
 
 then the monograph conditions supersede Ion Labs standard conditions.
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 D-725 6 Page 3 of 6 
 Loss on Drying 
 
 63.1 Equilibrate the oven to 105°C or appropriate temperature. Use a calibrated
 digital thermometer to confirm temperature of oven. 
 
 6.3.2 If the melting point of the sample is below 105°C then adjust temperature 5 to
 10°C below the melting point. 
 
 63.3 Dry a clean watch glass or Weigh Bottle before adding sample.
 
 6.3.4 Weigh a clean and dry watch glass or weigh bottle on the analytical balance.
 
 6.3.5 Print the tare weight of the watch glass or weigh bottle. 
 
 6.3.6 Mix the substance to be tested and, if it is in the form of large particles, reduce
 
 the particle size to about 2 mm by crushing before weighing. Apply a 1 to 2 g
 
 test sample to the tared container. 
 
 6.3.7 By gentle, sidewise shaking, distribute the test specimen as evenly as practicable
 
 to a depth of about 5 mm generally, and NMT 10 mm in the case of low bulk
 density materials. 
 
 6.3.8 Print the sample weight. 
 
 6.3.9 The weight of the tare and sample is the initial weight. 
 
 6.3.10 Place the open container holding sample into a pre-equilibrated oven.
 
 6.3.11 If necessary, apply 20 +/- 4 inches of vacuum and dry for 2 hours. Once
 
 vacuum is applied, ensure oven temperature is at testing conditions.
 
 6.3.12 After allotted time, be sure to cap samples to keep them from drawing moisture
 
 from the air and move the sample to a desiccator to cool for 20 minutes or until
 
 the sample has equilibrated to room temperature. Use caution when moving
 samples as they may be hot. Weigh the sample again after cooling. This is the
 
 final weight of container and sample. 
 
 6.3.13 Print the final container and sample weight. 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-725 Page 4 of 6 
 Loss on Drying 
 
 6.3.14 The % LOD is determined by the following equation: 
 
 initial weight — final weight 
 %LOD = x 100% 
 sample weight 
 
 6.3.15 Record results on Form D-725-F1 LOD Test Record. 
 
 6.4 Cautionary Measures when using a drying oven 
 
 6.4.1 Cautionary measures should be taken when drying multiple samples
 
 concurrently. 
 
 6.4.2 A recommended maximum of 10 samples should be dried in the oven at the
 
 same time. 
 
 6.4.3 Care should be taken when drying samples with higher moisture content with
 
 those containing lower moisture contents and hygroscopic in nature.
 
 6.5 Testing - Halogen Moisture Analyzer 
 
 6.5.1 Monthly Check 
 
 6.5.1.1 Level Instrument. 
 
 6.5.1.2 Place empty sample pan into the sample pan handler. Place sample
 
 pan handler into the draft shield, ensuring the tongue of the sample pan
 handler lies exactly in the slot of the draft shield. The pan must lie flat
 
 in the pan holder. 
 
 6.5.1.3 Close lid. The instrument will tare automatically. 
 
 6.5.1.4 Remove a SmartCal stick from blister pack, open and distribute
 contents evenly over the sample pan. Close lid. 
 
 6.5.1.5 Press menu and select “CAL”. In “CAL” menu, select “S.CAL”.
 
 6.5.1.6 Set temperature that closely matches the preferred application.
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-725 Page 5 of 6 
 Loss on Drying 
 
 Confirm entry with the backwards arrow button. 
 
 6.5.1.7 Compare the test results to the below table and evaluate whether the
 
 result meets the control limits 
 
 3} ¥ x ‘A =|f Steeys)h ases teesaaoee Fs $,. g|E RTESSE Je?BSS* ee SSSPh ESAGSSs
 ‘ a c T S. A i aE 4’ RIZRAIE J SS NS: 5 es x5 sy Re : Measnasaneens ei
 2-4.4% 
 ° 
 5.2 — 6.4% 
 7.4 — 8.8% 
 9.9 —11.7% 
 6.5.1.8 Capture monthly check in instrument logbook/ notebook.
 6.5.1.9 If the test falls outside of the above control limits, take the following
 
 measures: 
 
 6.5.1.9.1 After instrument has cooled down, repeat the test and
 
 ensure all steps have been carried out correctly.
 
 6.5.1.9.2 If the failure persists, perform a temperature and weight
 
 adjustment and perform an additional SmartCal test.
 
 6.5.1.9.3 If the result continues to fail, reference Troubleshooting
 
 guide in instrument manual. 
 
 6.5.2 Performing a simple measurement 
 
 6.5.2.1 Let instrument warm up for one hour. 
 
 6.5.2.2 Place empty sample pan into the sample pan handler. Place sample
 pan handler into the draft shield, ensuring the tongue of the sample pan
 
 handler lies exactly in the slot of the draft shield. The pan must lie flat
 in the pan holder. 
 
 6.5.2.3 Close lid. The instrument will tare automatically. 
 
 6.5.2.4 Open lid and place specimen into sample pan. A minimum of 0.5
 
 grams of sample is required. Ensure the sample is distributed evenly to
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-725 Page 6 of 6 
 Loss on Drying 
 
 obtain good analysis results. 
 
 6.5.2.5 Close the heating module lid. The drying and measuring process starts
 
 automatically. 
 
 6.5.2.6 Results will automatically print. 
 
 6.5.2.7. Capture results on Form D-725-F2 Halogen Moisture Analyzer Test
 Record. 
 
 7.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 10/05/10 | New procedure. ~ - l 01/26/12 Updated SOP format - - | - | - |
| 2 | 02/21/13 | Reformatted = ‘ Two forms added, updated instrument and supply information, more | - | - |
| 3 | 08/28/13 | detailed instruction on use of the vacuum oven and electronic moisture analyzer, removed USP <921> references Biennial review: Reformatted SOP. Removed electronic moisture | 13-764 | B. Johns |
| 4 | 01/05/16 | analyzer instructions and form. Added use of temperature probe. Added tolerances. | 16-0023 | N. Zhang |
| 5 | 02/04/19 | peak review: updated responsibilities, fixed typos, clarified | 19-0117 | J. Maipnan |
| 6 | 06/06/22 | Update to reflect current practices. Added additional instruction to CC- incorporate Halogen Moisture analyzer. Updated logo and format. | 22-0159 | J. Sassman |

 

[SOP 

 IANLABS LOD Test Record 
 Form: D-725-F 1 CCR No. CC-22-0159 Revision: 4 
 
 (Initial Weight - Final Weight) X 100 
 Sample Weight 
 
 Determination (circle one): Pass Fail 
 Comments: 
 
 Performed By/Date: 
 
 Reviewed By/Date: 
 
 

[SOP 

 IANLABS Halogen Moisture Analyzer Test Record 
 Form: D-725-F2 CCR No. CC-22-0159 Revision: 0 
 
 Analyzer Result Tape: 
 
 Determination (circle one): Pass Fail 
 Comments: 
 
 Performed By/Date: 
 
 Reviewed By/Date: