D-744

Lowry Protein Determination

Section D — Laboratory Operations and Specifications Revision 3 12 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 The purpose of this procedure is to describe a method for the quantitative analysis of protein
 content in raw materials and finished products using the Protein Lowry method. The method is
 
 based on the USP monograph <1057> Biotechnology-Derived Articles- Total Protein Assay,
 
 method 2. 
 
 2.0 Scope 
 
 The principle of the Protein Lowry method involves the reaction of protein, cupric sulfate and
 tartrate in alkaline solution, resulting in formation of tetradentate copper-protein complexes.
 
 When Folin-Ciocalteu Reagent is added, it is effectively reduced in proportion to these chelated
 copper complexes, producing a water-soluble product whose blue color can be measured at
 
 750nm. Some excipients and dietary ingredients used in finished products can interfere with
 
 the analysis of protein thus each finished product containing protein with a label claim needs to
 be verified against interference and/or suppression before using the method. The methods are
 
 structured based on the monograph <1057> Biotechnology-Derived Articles- Total Protein
 Assay, method 2. This method was validated for use with dietary supplements under protocol
 
 MV-LAB-14-013. 
 
 3.0 Responsibility 
 
 3.1 It is the responsibility of QC and Analytical Chemists to follow this procedure.
 
 3.2 It is the responsibility of QC Laboratory Management to implement this procedure and
 to ensure that the procedure is being followed. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-744 Page 2 of 12 
 Lowry Protein Determination 
 
 3.3 It is the responsibility of QC Laboratory Management and/or Analytical Development
 to keep this procedure aligned with current practices. 
 
 4.0 Definitions 
 
 4.1 STD — Standard 
 
 4.2 SAM — Sample 
 
 4.3. CofA — Certificate of Analysis 
 
 4.4 NaOH — Sodium Hydroxide 
 
 4.5 NaCl -— Sodium Chloride 
 
 4.6 SDS -— Sodium Dodecyl Sulfate 
 
 4.7 QC — Quality Control 
 
 5.0 References 
 
 5.1 USP monograph <1057> Biotechnology-Derived Articles- Total Protein Assay,
 
 method-2 
 
 5.2. MV-LAB-14-013, Protocol, Total Protein Determination using the Protein Lowry and
 
 Bradford methods coupled with Visible Light Spectroscopy 
 
 6.0 Reagents, Supplies and Equipment 
 
 6.1 Reagents: general- reagent grade or better 
 
 6.1.1 Millipore water 
 
 6.1.2 Sodium Chloride (NaCl) 
 
 6.1.3. Sodium Hydroxide (NaOH) 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-744 Page 3 of 12 
 Lowry Protein Determination 
 
 6.1.4 Sodium Dodecyl] Sulfate (SDS) 
 
 6.1.5 Lowry Reagents- Formulated as a kit by Thermo Scientific, Cat# 23240 or
 
 equivalent 
 
 6.1.6 Protein Standard(s) 
 
 6.2 Supplies and Glassware 
 
 6.2.1 13mm X 100mm glass test tubes with caps 
 
 6.2.2 Scintillation vials 
 
 6.2.3. 25ml, 50ml, and 100m1 volumetric flask 
 
 6.2.4 200uL, 1ml and 10ml pipette tips 
 
 6.2.5. 1.5ml and 2.0m] Eppendorf tubes 
 
 6.2.6 Plastic luer lock syringes 
 
 6.2.7 Weigh boats 
 
 6.2.8 Nylon syringe filters, 0.45um 
 
 6.2.9 Whatman #1 filter paper 
 
 6.2.10 Disposable Cuvettes- regular or semi-micro 
 
 6.3 Equipment 
 
 6.3.1 Lambda 45 Spectrophotometer, Perkin Elmer, with Win-Lab Software or
 
 equivalent 
 
 6.3.2 Analytical Balance 
 
 6.3.3. Vortex 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 D-744 3 Page 4 of 12 
 Lowry Protein Determination 
 
 6.3.4 Wrist-Action Shaker 
 
 6.3.5 Eppendorf Centrifuge 
 
 6.3.6 200ul, 1ml and 10ml1 pipettes 
 
 6.4 Buffers 
 
 6.4.1 Lowry Dissolution Buffer (1)- 0.9% NaCl, 0.05M NaOH, 0.1% SDS in
 
 Millipore water, prepared by adding 9.0g NaCl, 1.0g SDS and 2.0g NaOH to a
 1L volumetric flask and dissolving to 1L with Millipore water. Other final
 
 amounts can be made by adjusting all ratios to meet that final volume.
 
 6.4.2 Lowry Dissolution Buffer (2) <for egg protein> 0.9% NaCl and 0.1%SDS in
 
 Millipore water, prepared by adding 9.0g NaCl and 1.0g SDS to a 1L volumetric
 flask and dissolving to 1L with Millipore water. Other final amounts can be
 
 made by adjusting all ratios to meet that final volume. 
 
 7.0 Sample and Standard preparation 
 
 7.1 Standard Certification 
 
 7.1.1 For raw materials that do not have a protein standard that is commercially
 available, a raw material consisting of a similar protein composition can be
 
 certified for use as a standard by determining the protein content using nitrogen
 
 pyrolysis or other appropriate source on a dry basis providing no other
 appreciable nitrogen sources are contained within the material.
 
 7.1.2 Finished product protein standards will be generated by mixing in proportion to
 the formula all protein types. The finished product protein blend will be
 
 certified for use as a standard by determining the protein content using nitrogen
 pyrolysis on a dry basis providing no other appreciable nitrogen sources are
 
 contained within the blend. 
 
 7.1.3 If, during the validation of the standard a slope is generated in the linear range
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 D-744 3 Page 5 of 12 
 Lowry Protein Determination 
 
 that is equivalent to another standard, the standards with matching slopes can be
 used interchangeably. 
 
 7.1.4 Protein standards are certified yearly. 
 
 7.2 Standard Preparation 
 
 7.2.1 Use the tested purity value from nitrogen pyrolysis on a dry basis or the CofA
 
 for the specific protein standard in your calculations. The Stock Standard
 Preparation reflects 100% content. 
 
 Example: Egg White, 81.3% protein standard 
 
 Prepare 50 ml of a Img/ml solution 
 
 Img/ml X 50ml = 50mg 
 
 50mg/0.813 = 61.5mg 
 
 Dissolve 61.5mg up to 50m1 = 1.0mg/ml 
 
 7.2.2 Dry approximately one gram of standard material under vacuum at 105°C for 2
 
 hours using a scintillation vial. Allow material to cool before weighing.
 
 7.2.3 All Standards are prepared by weighing no less than the limit of the scale used,
 
 during that time period and transferring to a volumetric flask of the appropriate
 
 volume. Bring up to 2/3’s final volume using the appropriate Dissolution Buffer.
 Vortex for 1 minute to facilitate dissolution. Attach the flask to a wrist action
 
 shaker and mix for an additional 15 minutes. Bring standard to final volume
 before using. 
 
 7.2.4 Begin with a 1 mg/ml standard stock. Dilute the standard to a concentration
 reflective of the estimated protein concentration of the sample using the
 
 appropriate Dissolution Buffer. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 D-744 3 Page 6 of 12 
 Lowry Protein Determination 
 
 7.3 Sample Preparation 
 
 7.3.1 If samples are being reported on a dry basis, dry approximately one gram of
 
 sample material under vacuum at 105°C for 2 hours in a scintillation vial. Allow
 
 material to cool before weighing. 
 
 7.3.2 No less than the limit of the scale used, during that time period of material may
 
 be weighed. 
 
 7.3.3 Weigh an exact amount of material (not corrected for purity) to make a 1 mg/ml
 
 sample. 
 
 @ Other concentrations can be prepared to address solubility or other related
 issues. 
 
 7.3.4 Samples can be dissolved in Dissolution Buffer at any volume starting from
 25mL. 
 
 7.3.5 All samples are shaken on a wrist action shaker for 15 minutes at RT in two
 thirds their initial volumes then brought up to final volume.
 
 7.3.6 Typical high protein content samples should be diluted to 0.75mg/ml using
 Dissolution Buffer for testing. Other concentrations can be used.
 
 7.3.7 The final diluted sample must be filtered or centrifuged before analyzing.
 
 7.3.8 For finished products or raw materials being analyzed for the first time using
 
 either method an in process verification is required to demonstrate the absence
 
 of interference and/or suppression before quantitation can be implemented.
 
 8.0 Lowry Method for Identification 
 
 8.1 Prepare 1X (1N) Folin-Ciocalteu Reagent by diluting supplied 2X (2N) reagent 1:1 with
 
 Millipore water by adding 500ul 2X Reagent and 500ul Millipore water to a 2ml
 Eppendorf tube and vortex. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev Pave 7 of 12 
 Lowry Protein Determination D-744 3 8 
 
 8.2 Add solution to test tubes as follows: 
 
 8.2.1 Add 300ul of each sample or standard into an appropriately labeled test tube.
 
 8.2.2 Add 1.5ml of Modified Lowry Reagent to each test tube. Mix well and incubate
 each tube at room temperature for exactly 10 minutes. 
 
 8.2.3 At the end of the 10-minute incubation period, add 150ul of prepared 1X Folin-
 
 Ciocalteu Reagent and vortex to mix the contents. 
 
 8.2.4 If a color change is not immediately observed, cover sample mixture and
 
 incubate up to 30 minutes. 
 
 8.2.5 A color change to blue or purple is acceptable as passing unless otherwise
 
 specified. 
 
 9.0 Lowry Method for Quantitation 
 
 9.1 Parameters 
 
 9.1.1 Wavelength= 750nm 
 
 9.1.2 Slit width- 2mm 
 
 9.1.3 Range- lug to 1.5mg 
 
 9.2 See Attachment I for dilution schedule 
 
 9.3. Prepare 1X (1N) Folin-Ciocalteu Reagent by diluting supplied 2X (2N) reagent 1:1 with
 Millipore water by adding 500ul 2X Reagent and 500ul Millipore water to a 2ml
 
 Eppendorf tube and vortex. 
 
 9.4 Add solution to test tubes as follows: 
 
 9.4.1 Add 300ul of each sample or standard into an appropriately labeled test tube.
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-744 Page 8 of 12 
 Lowry Protein Determination 
 
 9.4.2 At 15-second intervals, add 1.5ml of Modified Lowry Reagent to each test tube.
 Mix well and incubate each tube at room temperature for exactly 10 minutes.
 
 9.4.3 Exactly at the end of the 10-minute incubation period, add 150u1 of prepared 1X
 
 Folin-Ciocalteu Reagent, immediately vortex to mix the contents.
 
 9.4.4 Cover (completely eliminate light exposure) and incubate all tubes at RT for 30
 
 minutes. 
 
 9.4.5 Transfer samples into cuvettes. 
 
 9.4.6 Measure absorbance at 750nm. 
 
 10.0 Data Collection 
 
 9.1 Turn on the Lambda 45 UV/VIS Spectrophotometer and allow the bulbs to warm up for
 
 at least 15 minutes. 
 
 9.2 Open <WinLab> Software. 
 
 9.3 Click on <Methods>. 
 
 9.4 Open task folder. 
 
 9.5 Click on <D-744> Test Method. 
 
 9.6 Click on <Run>. 
 
 9.7 Choose the number of samples. 
 
 9.8 Complete sample spreadsheet to include: 
 
 9.8.1 Sample ID 
 
 9.8.2 Description 
 
 e Type — sample, standard, blank, control 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-744 Page 9 of 12 
 Lowry Protein Determination 
 
 e Concentration 
 
 9.8.3 Press <Start> to initiate testing. 
 
 9.8.4 Press <OK> to Zero the spectrophotometer. 
 
 9.8.5 Place the cuvette(s) in the designated slot(s) on the carousel.
 
 9.8.6 Press <OK> to start data Collection 
 
 9.8.7 The results can be previewed by clicking on the <Results> icon.
 
 9.8.8 Save as <New Task>. 
 
 e Label file using Date-Sample ID-Concentration format. 
 
 9.8.9 Click on the <Output> icon. 
 
 9.8.10 Select <Default-Wavelength Program (Rev1)> from the drop down menu.
 
 9.8.11 Print. 
 
 9.9 Calculations for determining quantity 
 
 9.9.1 The protein concentration in a single dose can be calculated by the following
 
 equation when a single point standard is used: 
 
 e %wt = ((ACABS7s0sam / ABS7so0sta))/B) 
 
 A = Standard concentration in mg/mL 
 
 B = Sample concentration in mg/mL 
 
 ABS750sam = Average absorbance of sample after blank subtraction
 
 ABS7s0sta= Average absorbance of standard after blank subtraction
 
 % wt = The percent protein in the sample 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev Pave 10 of 12 
 Lowry Protein Determination D-744 3 g 
 
 9.9.2 The protein concentration in the sample can be calculated by the following
 equation when a multipoint standard curve is generated: 
 
 @ Equation 1: Y = mX, linear regression forced through 0,0 
 
 Equation 2: Ywt = mX/B 
 
 Y = Protein concentration of sample in mg/ml 
 
 B = Sample concentration in mg/ml 
 
 % wt = The percent protein in the sample 
 
 m = slope of standard curve 
 
 X = Absorbance of sample average after blank subtraction 
 
 9.9.3 Calculation for determining protein content per serving 
 
 e@ C= Swt* wtaose 
 
 C = weight of protein in a dose 
 
 % wt = The percent protein in the dose 
 
 Wldose = The weight of a single dose 
 
 9.9.4 The % of label in a single dose of protein can be calculated by the following
 equation: 
 
 e % of label =C\LC 
 
 C = weight of protein in a dose 
 
 LC = Label content for a single dose 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Lowry Protei° n Determi.n ati. on D-- 744 3 Page 11 of 12 
 
 11.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 05/17/14 | New | 14-0417 | B. Johns |
| 1 | 09/21/16 | Biennial review: Updated SOP format. Updated stability and release requirements. Aligned SOP with current practices, Removed release CC-20- . | 16-0850 | N. Zhang |
| 7 | 04/14/20 | requirements as it is in test tickets. 0274 J. Maignan | - | - |
| 3 | 08/08/24 | Split out section 8.0 to differentiate between identification and CC-24- A. Shannon quantitation of samples. 0366 | - | - |

 12.0 Attachments 
 
 12.1. Attachment | — Standard Dilution Schedule for Sample Tubes 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 D-744 Page 12 of 12 
 Lowry Protein Determination 
 
 Attachment 1 — Standard Dilution Schedule for Sample Tubes 
 
 1 2.0 
 2 1.0 2.0m¢/ml 300 300 
 3 0.50 1.0 mg/ml 300 300 
 4 0.75 1.0 mg/ml 450 150 
 5 0.25 0.5 mg/ml 300 300 
 6 0.05 0.25 mg/ml 120 A480 
 
 1 1.4 140 140 60 2.0mg/ml stock 
 
 2 1.0 100 100 100 2.0mg/ml stock 
 3 0.75 75 75 125 2.0mg/ml stock 
 4 0.50 50 50 150 2.0mg/ml stock 
 5 025 25 50 150 1.0 mg/ml std 
 6 0.150 15 30 170 1.0 mg/ml std 
 7 0.125 12.5 50 150 0.5 mg/ml std 
 8 0.100 10 AO 160 0.5 mg/ml std 
 9 0.050 5 40 160 0.25 mg/ml std 
 10 0.010 1 AO 160 0.05 mg/ml std