D-805

Calibration Verification and Operation of the AquaLab 4TE

Section D — Laboratory Operations and Specifications Revision 1 5 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 This procedure provides instruction for the calibration, verification and operation of a water
 activity meter and specific instruction for the operation of the AquaLab 4TE.
 2.0 Scope 
 
 This procedure applies to the AquaLab 4TE water activity meter in use at Ion Labs, Inc.
 
 3.0 Responsibility 
 3) It is the responsibility of the QC Laboratory personnel to comply with this SOP.
 
 3.2 The QC Laboratory management is responsible for ensuring compliance with this SOP.
 3.3. It is the responsibility of the QC Laboratory/Analytical Development management to
 keep this SOP current with the latest Ion Labs practices. 
 
 4.0 Definitions 
 4.1 QC - Quality Control 
 
 4.2 SOP — Standard Operating Procedure 
 4.3 K — Thousand 
 
 4.4 M-Million 
 
 4.5 Full Scale Range — A calculation of the maximum viscosity value that can be measured
 by a spindle and speed combination when used with the specific spring torque of the
 DV1 viscometer 
 5.0 References 
 
 5.1 AquaLab Series 4 Water activity meter Quickstart Guide 
 5.2. AquaLab Series 4 Water activity meter Standard Operating Procedure
 
 5.3 D-805-F1, Form, Water Activity Measurement Ticket 
 6.0 Procedure 
 
 6.1 Location and Setup 
 
 6.1.1 The water activity meter requires a clean, level surface in a temperature-stable
 environment for proper operation. 
 6.1.2 The meter is powered up by first plugging in then flipping the rocker switch on
 the back of the instrument. Allow the unit to warm up for 15 minutes.
 
 
 

[SOP 

 Standard Operating Procedure SOPNo | Rev Page 
 Calibration, Verification and Operation of the D-805 1 2 of5 
 AquaLab 4TE 
 
 6.2 Performance Verification 
 
 6.2.1 Verification should be conducted at least once per day before use or if readings
 become unstable. 
 
 This flowchart is a graphical representation of the directions given below for performance verification.
 
 7 
 aon 
 4d Measure Calibration Standard \ 
 Repent Process 4 
 
 Clean Sample Chamber Met Correct Comect 
 Clean Sample Chamber 
 Het Correct = Not Comact 
 Re-Read Standard —_— Measure 2° Standard 
 
 Not Correct 
 Correct 
 = 
 
 Go to Linear Offset Procedure Go to Sampling Procedure ; 
 / 
 
 6.2.2 Verify the performance of the instrument with calibration standards that have
 known water activities: 
 
 Calibration Standard @25C | Molalitv | Water Activity 
 Distilled Water 1.000 20.003 
 KCL 05m 0.984 £0.003 
 NaCl 6.0m 0.760 20.003 
 Lidl 8.57 m 0.50020.003 
 Licl Udim | 025020008 
 6.2.3 Choose at least two verification standards that are on each side of the
 approximate samples you would like to measure. 
 6.2.4 Empty a vial of the calibration standard solution into a sample cup and place in
 the sample chamber. 
 
 6.2.5 Close lid and move the lever to the READ position. 
 6.2.6 The water activity reading of each standard should be within +0.003 ay of its
 
 specification. 
 6.2.7 If your first standard is not within +0.003 ay of its specification, it is probably
 due to contamination of the sample chamber. Refer to section 7.0 for cleaning
 instructions, 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev Page 
 Calibration, Verification and Operation of the D-805 1 3 of5 
 AquaLab 4TE 
 
 6.2.8 After cleaning, repeat verification. If reading are consistently outside of the aw
 range a linear offset may have occurred and the reading of the standard should
 be corrected. Refer to section 6.3 for Linear offset procedure instructions.
 6.2.9 If both readings are within the aw range, proceed to section 6.4 for sample
 running procedure. 
 
 6.3 Linear Offset Procedure 
 6.3.1 When there is confidence that a linear offset has occurred, toggle to the
 
 Configuration tab by pressing the Menu icon button. Calibration is the first
 option highlighted in the configuration tab. Press the Enter icon button to access
 the calibration routine. Follow the onscreen instructions to perform the linear
 offset. They are listed below for convenience. 
 6.3.2 Press the Enter button to start the linear offset process. To return to the main
 menu, press the cancel button. Follow the prompt to insert a fresh standard and
 seal the chamber. 
 
 6.3.2.1 The same sample that triggered the need for the linear offset can be
 used to adjust the linear offset. Be sure that the rim and the outside of
 the cup are clean. 
 6.3.3 Close the lid and slide the lever to the READ position. Press the CHECK icon to
 start the test. You can cancel the linear offset program by returning the lever to
 
 the OPEN position or pressing the (X) button to cancel. 
 6.3.4 After the calibration standard is complete, press the up and down arrows to
 adjust the ay reading to its proper value for the standard being measured.
 
 6:35 When the value is correct, press the Save icon button to store the new value. To
 return to the main menu, press the Cancel button and no changes will be made.
 6.3.6 Re-test the calibration standard in normal sampling mode (Section 6.2.5). If the
 standard is within the proper accepted range, continue with the second standard.
 
 6.4 Sample Testing 
 6.4.1 It is important to ensure the sample should be representative of the product.
 
 6.4.2 Be sure the completely cover the bottom of a clean, disposable sample cup.
 6.4.3 To minimize the chance to contaminate the sample chamber, be sure the sample
 cup is not more than half full. 
 
 6.4.3.1 If testing gummy product forms, using a razor or knife, thinly slice a
 gummy and layer into a sample cup. 
 
 *Note: Be very careful when using a razor or knife to cut gummies*
 6.4.4 Place sample cup, removing any lids on the sample cup, in a clean chamber and
 close the lid of the system. 
 6.4.5 Move the lever to the READ position. 
 
 6.4.6 After samples have finishing measuring, be sure to not leave samples in the
 chamber for extended periods of time. 
 
 
 

[SOP 

 Standard Operating Procedure SOPNo | Rev Page 
 Calibration, Verification and Operation of the D-805 1 4 of5 
 AquaLab 4TE 
 
 6.4.7 Record the ay of the sample on appropriate form. 
 
 6.5 Calibration Notebook and Form D-805-F1 
 
 6.5.1 This notebook will record the following information each time the water activity
 meter is calibrated 
 6.5.1.1 Analyst initials 
 
 6.5.1.2 Date 
 6.5.1.3 Standard theoretical value and obtained value (should be +0.003aw)
 
 6.5.2 Form D-805-F1 can be used for raw material and finished product testing and
 should accompany the Internal Certificate of Analysis. 
 6.5.3 Results can also be documented in a laboratory notebook and a copy can be made
 to accompany the Internal Certificate of Analysis. 
 
 7.0 Cleaning 
 
 7.1 Keeping the instrument clean is vital to maintaining the accuracy of the instrument.
 Dust and debris from samples, oils from fingerprints on the mirror, or other sources can
 contaminate the chamber. The mirror must be perfectly clean for a smooth and even
 dew formation. A visual inspection should be performed often and at least as part of the
 verification process. A cleaning video is available at 
 metergroup.com/meter_knowledgebase/aqualab-4te-certification 
 
 THGPTIOBILE op 
 Capacitance sensor ‘ ? 
 filter (TEV only} 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev Page 
 Calibration, Verification and Operation of the D-805 5 of 5 
 AquaLab 4TE 
 
 7.2 Be sure to wear clean gloves when cleaning the instrument. 
 7.3 With the instrument off, open the chamber to expose the sample chamber and sensors.
 
 7.4 Clean the sample chamber (any area inside of the red O-ring with a closed lid) with
 isopropyl alcohol and Kimwipes or other appropriate cleaning solutions.
 7.5 Be extremely careful around the fan blades when cleaning the chamber.
 
 7.6 Remove any debris from the around or within the sample chamber. 
 7.7 You may need to wrap a moistened Kimwipe around the end of a spatula and gently
 clean hard to reach areas. 
 
 7.8 Repeat with deionized water and after with dry Kimwipes to remove any moisture.
 7.9 After cleaning, visually inspect the chamber and sensers for any contaminant (including
 moisture) and let stand for at least 5 min for ensure chamber is dry.
 
 7.10 After cleaning, be sure to perform a system verification and perform any correct for any
 linear offset that may have occurred. 
 8.0 Water Activities vs Growth of Representative Microorganisms 
 
 8.1 Per USP 1112 water activity can be used to justify reduced microbial testing.
 
 Bacteria Water Activity (aw) Molds and Yeast Water Activity (aw) 
 P. aerginosa 0.97 R. nigricans 0.93 
 B. cereus 0.95 M. plumbeus 0.92 
 C. botulinum, TypeA 0.95 R. mucilaginosa 0.92 
 E. coli 0.95 S. cerevisiae 0.90 
 C. perfringens 0.95 P. variotti 0.84 
 L. viridescens 0.95 P. chrysogenum 0.83 
 S. spp. 0.95 A. fumigatus 0.82 
 E. aerogenes 0.94 P. glabrum 0.81 
 
 B. subtilis 0.90 A. Flavus 0.78 
 M. lysodekitcus 0.93 A. niger 0.77 
 S. aureus 0.86 Z. rouxii 0.62 
 | halobium 0.75 X. bisporus 0.61 
 
 9.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 01/14/19 | New. N/A J. Maignan | - | - |
| 1 | 03/06/20 | Add instructions for testing gummies CC- | 20-0180 | S. Sassman |