D-808

Use of OpenLab for HPLC and GC Data Acquisition and Reporting

Section D — Laboratory Operations and Specifications Revision 2 19 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 This procedure provides guidelines for the use of Agilent OpenLab Chromatography Data
 
 System for HPLC instrument control, data processing, and reporting. 
 
 2.0 Scope 
 
 This procedure applies to all chromatographic data acquired using OpenLab Chromatography
 
 Data System in the Quality Control laboratory at Ion Labs. 
 
 3.0 Responsibility 
 
 | It is the responsibility of QC Laboratory analysts to follow this procedure.
 
 3.2 QC Laboratory Management and/or Analytical Development personnel are responsible
 
 for ensuring analysts follow the guidelines set forth herein. 
 
 3.3 It is the responsibility of QC Laboratory Management and/or Analytical Development
 personnel to keep this procedure aligned with current practices. 
 
 4.0 Definitions 
 
 4.1 QC — Quality Control 
 
 4.2 CDS — Chromatography Data System 
 
 4.3 GC — Gas Chromatography 
 
 4.4 HPLC — High Performance Liquid Chromatography 
 
 5.0 References 
 
 5.1 D-808-F1, Form, HPLC/ GC Analyst and Data Reviewer Checklist 
 
 3.2 D-807, SOP, HPLC Operation Maintenance and Qualification 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition D-808 2 | Page 2 of 18 
 
 and Reporting 
 
 5.3 D-801, SOP, Agilent 7890 GC Operation Maintenance and Qualification
 
 5.4 D-903, SOP, Conversion Factors Used in Analytical Determinations and New Product
 
 Formulation 
 
 6.0 General Documentation and Reporting Guidelines 
 
 6.1 The following parameters should be recorded in the laboratory notebook:
 
 6.1.1 Instrument Ion Number 
 6.1.2 Instrument Cal Due Date 
 
 6.1.3 Column identifier 
 
 6.1.4 Sequence Identification 
 6.2 HPLC/ GC Data Review Checklist 
 
 6.2.1 Each sequence generated from the use of HPLC and/ or GC should include Form
 D-808-F1 HPLC/ GC Analyst and Reviewer Checklist. 
 
 6.2.1.1 In the event a sequence was rejected for any reason (e.g. system suit
 failure), this checklist is still required to be completed, listing the
 rationale for the rejection. 
 
 6.2.2 This checklist is completed by both the analyst and the data reviewer to ensure all
 essential information was captured and is present in the data packet.
 
 6.2.3 Additionally, both the analyst and the data reviewer use the checklist to show
 assessment of all method suitability requirements. 
 
 7.0 Data Acquisition 
 
 TA If not already started, start the OpenLab Control Panel from the Windows start menu by
 selecting Agilent Technologies — Control Panel. Alternatively, click the Control Panel
 
 shortcut on the desktop. 
 
 72 In the bottom left corner of the Control Panel, select Instruments.
 
 7.3 On the left side under Instruments, select the desired instrument.
 
 74 Select the desired Project near the center/top of the screen. 
 
 7S Click Launch on the right side of the screen. 
 
 7.6 The data acquisition window opens. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | -808 2 | Page 3 of 18 
 
 and Reporting 
 
 7.7 To create a HPLC data acquisition method: 
 
 7.7.1 Click the Method Icon fa) at the top of the screen. 
 
 7.7.2 Click the New Icon “at the top left of the main window. 
 
 7.7.3 Under Instrument Setup — Pump, set the following to match the analyte specific
 method: 
 
 7.7.3.1 Flow rate. 
 
 7.7.3.2 Tnitial solvent makeup. 
 
 7.73.3 Gradient timetable. 
 
 7.73.4 Stoptime. 
 
 7.7.4 Under Instrument Setup > Sampler: 
 
 7.74.1 Set the injection volume to match the analyte specific method.
 
 7.74.2 Set stoptime to “As Pump/No Limit”. 
 
 7.7.5 Under Instrument Setup — Column Comp: 
 
 7.7.5.1 Set column temperature to match the method. 
 
 7.7.5.2 Set Enable Analysis when temperature is within + 0.8 °C.
 
 7.7.5.3, Set stoptime to “As Pump/Injector”. 
 
 7.7.6 Under Instrument Setup — DAD: 
 
 7.7.6.1 Set the wavelength for each channel required by the analyte specific
 
 method. 
 
 7.7.6.2 Set the bandwidth for each channel to 4 nm (or that required by the
 
 analyte specific method). 
 
 7.7.6.3 Ifusing a reference, set the reference wavelength(s) and bandwidth(s).
 
 7.7.6.4 Set the Peakwidth to an appropriate value (e.g. HPLC = 2.5 Hz, UPLC
 = 10 Hz). 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition D-808 2 | Page 4 of 18 
 
 and Reporting 
 
 7.7.6.5 Set Stoptime to “As Pump/Injector”. 
 
 7.7.6.6 Under Spectrum, set Store to “All”, set the Range to match the analyte
 
 specific method, set Step to 2.0 nm. 
 
 7.7.6.7 Set margin for negative absorbance to 100 mAU. 
 
 7.7.6.8 Set the slit width to 4 nm. 
 
 7.76.9 Under Autobalance, select Prerun. 
 
 7.7.6.10 Under Lamps on required for acquisition: select UV lamp and Vis lamp.
 
 7.8 Tocreate anew GC data acquisition method: 
 
 7.8.1 Click the Method Icon [ill at the top of the screen. 
 
 7.8.2 Click the New Icon {1 at the top left of the main window. 
 
 7.8.3 On the left side, select Instrument Setup — Agilent 7890B. 
 
 7.8.4 Under Inlets — SSL — Front or Back (depending on which inlet you are using):
 
 7.8.4.1 Select the Heater check box, and set the inlet temperature to match the
 
 method. 
 
 7.8.4.2 Select the Pressure check box. Do not change the setpoint.
 
 7.8.4.3 Select the Septum Purge Flow check box, and set the septum purge flow
 to match the method. If not specified in the method, use a septum purge
 
 flow of 3 mL/min. 
 
 78.4.4 Set the Inlet Mode to Split or Splitless to match the method.
 
 7.8.4.5 If using split injection, set the Split Ratio. 
 
 78.4.6 Ensure that the Gas Save checkbox is not selected. 
 
 7.8.5 Under Inlets + SSL — Front or Back (for the inlet you are NOT USING):
 
 78.5.1 Ensure that the Heater, Pressure, and Septum Purge Flow check boxes
 
 are not selected. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | D-808 2 | Page 5 of 18 
 
 and Reporting 
 
 7.8.6 Under Configuration + Columns: 
 
 7.8.6.1 Select Line 1. 
 
 7.8.6.2 Click Catalog, and enter the column length, diameter, film thickness,
 
 column type, maximum temperature, maximum program temperature,
 minimum temperature, and description. Then click OK. 
 
 7.8.6.3 Select the Inlet that the column will be connected to.
 
 78.6.4 Select the Outlet that the column will be connected to.
 
 7.8.6.5 Select Heated By: Oven. 
 
 7.8.7 At the left side, select Columns: 
 
 78.7.1 Select column #1. 
 
 7.8.7.2 Under Control Mode, ensure the On checkbox is selected.
 
 7.8.7.3 Set the control mode to match the method: Constant Pressure, Ramped
 
 Pressure, Constant Flow, or Ramped Flow. 
 
 78.74 Set the Flow (or Pressure depending on control mode) to match the
 
 method. 
 
 78.7.5 If using ramped flow or ramped pressure, set the time program to match
 the method. 
 
 7.8.8 At the left side, select Oven: 
 
 7.8.8.1 Ensure the Oven Temp On checkbox is selected. 
 
 7.8.8.2 Under the Oven Temp On checkbox, enter the initial oven temperature
 
 specified in the method. 
 
 78.8.3 Enter the Equilibration Time specified in the method, or use a default of
 
 1 min. 
 
 7.8.8.4 Set the maximum oven temperature to 20 °C above the maximum
 temperature of the method. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition D-808 2 | Page 6 of 18 
 
 and Reporting 
 
 7.8.8.5 Adjust the oven temperature program to match the method.
 
 7.8.8.6 Set Post Run to 100 °C. 
 
 7.8.8.7 Set Post Run Time to 0 min. 
 
 7.8.9 At the left side, select Detectors + FID — Back: 
 
 7.8.9.1 Ensure that the Heater, Air Flow, Fuel Flow, Makeup Flow, and Flame
 
 check boxes are selected. 
 
 7.8.9.2 Adjust the Heater, Air Flow, Fuel Flow, and Makeup Flow settings to
 match the method. 
 
 7.8.9.3 Set the Carrier Gas Flow Correction to “Column + Makeup = Constant”.
 
 7.8.10 At the left side, select Aux Heaters: 
 
 7.8.10.1 If using headspace injection, ensure the checkbox is selected and set the
 
 temperature to match the transfer line temperature specified in the
 
 method. 
 
 7.8.11 At the left side, select Signals: 
 
 7.8.11.1 Set the Signal Source to Back Signal (FID). 
 
 7.8.11.2 Set the Data Rate to a value that is appropriate for the peak width. The
 data system should generate at least 10 points across the width of the
 
 narrowest peak. 
 
 7.8.11.3 The check boxes for Zero and Save should be selected.
 
 7.8.12 At the left side, select Readiness: 
 
 7.8.12.1 Ensure that only the components being used are selected.
 
 7.8.13 Under Instrument Setup, select Agilent 7697A. 
 
 7.8.14 Select Temperatures. If using headspace injection, ensure all boxes are checked
 
 and the temperatures match the method. Otherwise, ensure all boxes are not
 checked. 
 
 
 

[SOP 

 Standard Operating Procedure SOPNo | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | D-808 2 | Page 7 of 18 
 
 and Reporting 
 
 7.8.15 If using headspace, select Times. Set the times to match the method.
 
 7.8.16 Ifusing headspace, select Vial and Loop. Set the parameters to match the method.
 
 7.8.17 Ifusing headspace, select Advanced Functions. Select Single Extraction and Vent
 
 Vial Pressure After Extraction. If settings for Purge Flow and Purge Time are
 specified in the method, set Post Injection Purge to Custom, and change the
 
 settings for Purge flow and Purge time to match the method. If no settings for
 Purge Flow and Purge Time are specified in the method, set Post Injection Purge
 
 to Default. 
 
 7.8.18 If using headspace, select Sequence Actions. Set the actions to the desired values.
 
 7.9 For HPLC, purge the mobile phase channels that will be used by the method:
 
 7.10 Purge the HPLC system using the sequence of solvents recommended for the column
 
 required by the analyte specific method or as outlined in D-807 HPLC Operation
 Maintenance and Qualification 
 
 7.10.1 Click the Method Icon (al at the top of the screen. 
 
 7.10.2 At the left side of the screen under Instrument Setup, click Pump.
 
 7.10.3 Ensure that the pump purge valve is in the correct position as required by the
 
 desired operation (closed = flow through column, open = flow to waste).
 
 7.10.4 Set the desired flow rate and solvent composition. 
 
 7.10.5 Click send to instrument “4s, 
 
 7.10.6 Click the Status Icon at the top of the screen. 
 
 7.10.7 In the Instrument Status window, hover over the Pump section and click the green
 
 On button to initiate flow. To stop flow, click the red Off button.
 
 7.10.8 Repeat steps 7.10.1 to 7.10.7 for each of the solvents recommended for the
 
 column required by the analyte specific method or as outlined in D-807 HPLC
 Operation Maintenance and Qualification. 
 
 
 

[SOP 

 Standard Operating Procedure SOPNo | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition D-808 2 Page 8 of 18 
 
 and Reporting 
 
 711 For GC, Install the column specified in the method by following the directions outlined
 
 in D-801 GC Operation Maintenance and Qualification. 
 
 7.12 Open the data acquisition method: 
 7.12.1 Click the Method Icon (al at the top of the screen. 
 
 7.12.2 Click Open Acquisition Method [at the upper left side of the screen.
 
 7.12.3 Select the desired method, click Open. 
 
 Click send to instrument cs to download the method to the instrument.
7.13 Click the Status Icon FE at the top of the screen.

 714 
 
 7AS For HPLC, hover over each instrument module (pump, column oven, detector) and click
 on the green On button. The instrument begins equilibrating at initial conditions.
 
 To monitor the baseline, click the Status Icon EE atthe top of the screen. If the baseline
7.16 is not visible, click Online Signals at the top of the screen.

 TAT Whena stable baseline is achieved, the run may be started. 
 
 7.18 Creating and Run a Sequence 
 
 7.18.1 Click on the Sequence Icon at the top of the screen. 
 
 7.18.2 For each injection, enter the vial number, acquisition method, and sample name.
 
 7.18.3 Use the Add, Insert, and Delete icons FE fShS~S to add or delete rows as
 
 required. 
 
 7.18.4 Enter the desired name for the result set in the Result Name box at the bottom left.
 
 7.18.5 To start the run, click Run in the bottom right corner to begin the run.
 
 7.18.6 After the sequence has been started, it may be paused or aborted if necessary:
 
 7.18.6.1 Click on the Status Icon E, 
 
 7.18.6.2 If the Run Queue is not visible, click Run Queue at the top of the screen.
 
 7.18.6.3 In the Run Queue window, select the running sequence.
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | -808 2 | Page 9 of 18 
 
 and Reporting 
 
 7.18.6.4 Click Pause or Stop run at the top of the Run Queue window.
 
 7.18.7 Afier the sequence has been started, it may be altered if necessary:
 
 7.18.7.1 Click on the Sequence Icon at the top of the screen.
 
 7.18.7.2 Just above the sequence table and to the left, click on Edit.
 
 7.18.7.3 Make any necessary changes. 
 
 7.18.7.4 Click Update at the bottom of the screen. 
 
 8.0 Data Processing 
 
 8.1 Open the Sequence: 
 
 8.1.1 If not already started, start the OpenLabs Control Panel from the Windows start
 menu by selecting Agilent Technologies — Control Panel. Alternatively, click
 
 the Control Panel shortcut on the desktop. 
 
 Click on Projects in the lower left corner. 
 
 In the project list on the left, select the project where your data resides.
 
 At the top, click the Start Data Analysis icon 
 
 The Data Analysis window opens. 
 
 8.2 Setting Up the Processing Method 
 
 8.2.1 If the Processing Method has already been created, proceed to Section 8.3.
 
 Click on the Open Method icon il, open one of the default processing methods
8.2.2 to use as a template. For HPLC data, select “3D UV

 Quantitative DefaultMethod”. For GC data, select “GC_LC 
 Quantitative _DefaultMethod”. 
 
 8.2.3 At the top right, click on Processing Method if it is not already selected.
 
 8.2.4 The Processing Method will be displayed in the Main Window.
 
 8.2.5 Under General — Properties: 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | D-808 2 | Page 10 of 18 
 
 and Reporting 
 
 8.2.5.1 Select the Global tab. 
 
 8.2.5.2 Choose the desired integrator (ChemStation or EZChrom).
 
 8.2.6 Under Integration Events — Standard: 
 
 8.2.6.1 Set the Slope Sensitivity (ChemStation) or Threshold (EZChrom).
 
 Lower values increase integrator sensitivity (for smaller peaks), while
 larger values decrease integrator sensitivity (for larger peaks). A value
 
 of 1 is a good starting point. 
 
 8.2.6.2 Set the Peak Width (ChemStation) or Width (EZChrom) to a value equal
 to or less than the width at the base of the narrowest peak.
 
 8.2.6.3 Set the Area Reject and Height Reject values (ChemStation) or
 
 Minimum Area value (EZChrom). Any peak that is smaller than these
 
 values will not be integrated. 
 
 8.2.6.4 Signal specific events may be set if desired. For example, if monitoring
 210 nm and 254 nm, different values of Slope Sensitivity and Peak
 
 Width may be set for each wavelength. 
 
 8.2.7 Under Processing Method — Compounds — Identification: 
 
 8.2.7.1 For each target analyte, enter the analyte name, associated signal,
 expected retention time and retention time window (both can be
 
 modified later), 
 
 8.2.8 Under Processing Method —> Compounds — Calibration: 
 
 8.2.8.1 Select the General tab and set the following parameters:
 
 8.2.8.1.1 External Standard (most common) or Internal Standard
 
 depending on the method type. 
 
 8.2.8.1.2 Number of levels is usually 1 unless multiple standard
 concentrations are being analyzed. 
 
 8.2.8.1.3. Curve calculation: From individual calibration points.
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | D808 | 2 | Page 11 of 18 
 
 and Reporting 
 
 8.2.8.1.4 RF definition: Response per amount. 
 
 8.2.8.1.5 Normalize to: Not selected. 
 
 8.2.8.1.6 Concentration calculation: Amount * Multipliers * Dilution
 
 Factor. 
 
 8.2.8.1.7 Calculate mass %: Selected. 
 
 8.2.9 Under the Calibration + Compound Table tab: 
 
 8.2.9.1 If using an internal standard method, select the “Is ISTD” box for the
 analyte that is the internal standard. Enter 1 for ISTD amount.
 
 8.2.9.2 For all other analytes, set the following parameters:
 
 8.2.9.2.1_ Amount unit: Leave blank. 
 
 8.2.9.2.2 Concentration unit: Leave blank. 
 
 8.2.9.2.3 Response: Area or Height. 
 
 8.2.9.2.4 Associated ISTD: If using internal standard, select the
 
 associated internal standard. 
 
 8.2.9.2.5 Mode: Curve (or Reference if using another analyte as the
 
 calibrator RRF). 
 
 8.2.9.2.6 Weighting method: None (unless weighting justified by
 data). 
 
 8.2.9.2.7 Manual factor: Leave blank. 
 
 8.2.9.2.8 Curve reference: Leave blank (or enter the reference
 
 calibrator is using another analyte as the calibrator RRF).
 
 8.2.9.2.9 Ref. correction: 1.0 (or enter the RRF if using another analyte
 
 as the calibrator). 
 8.2.9.2.10 Curve model: Linear (unless other model justified by data).
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | D-808 2 | Page 12 of 18 
 
 and Reporting 
 
 8.2.9.2.11 Origin: Force (or Ignore if multiple concentrations of
 
 standards are used to generate a curve). 
 
 8.2.9.2.12 Multiplier: 1. 
 
 8.2.9.2.13 Levels: Leave blank. 
 
 8.2.10 Under Processing Method — Compounds, select Spectra (HPLC only).
 
 8.2.10.1 On the UV confirmation tab, set the following: 
 
 8.2.10.1.1 UV confirmation match factor limit = 900 
 
 8.2.10.1.2 Lower wavelength = lower wavelength specified in the
 method 
 
 8.2.10.1.3 Upper wavelength = upper wavelength specified in the
 
 method 
 
 8.2.10.1.4 Subtract baseline spectra should normally be selected. In rare
 cases, the spectral match may be improved by turning
 
 subtract baseline spectra off. 
 
 8.2.11 Under Processing Method — System Suitability, select Properties and choose the
 
 System Suitability parameters that are required by the method:
 
 8.2.11.1 For Tailing, Symmetry Factor, and Resolution, select Column
 Performance — All peaks. Select the USP box. 
 
 8.2.11.2 For Signal to Noise, select Signal to Noise — All peaks, Pharmacopeias
 
 — USP, Noise Calculation — P2P, and Noise range > Fixed from Start
 to End min on current where Start to End is a range in the chromatogram
 
 where no peaks elute (noise range). 
 
 8.2.12 Under Processing Method — Tools: 
 
 8.2.12.1 Select Custom Calculation. 
 
 8.2.12,.2 Under Linked File, click Browse and select 
 
 “Calc_Percent_Label_Claim.ccf”. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | -808 2 | Page 13 of 18 
 
 and Reporting 
 
 8.2.13 Click the Save Method icon to save the processing method. 
 
 8.3 Setting Up the Injection List and Processing the Data. 
 
 8.3.1 Under Data Selection on the left side, select the sequence that contains your data
 
 and click Load Data at the top left. 
 
 8.3.2 The Data Processing window opens. 
 
 8.3.3 If the Create New Processing Method dialog box pops up, click No Method.
 
 8.3.4 At the top, click the Open Method icon fail, 
 
 8.3.5 Select the desired processing method, and click Open. 
 
 8.3.6 On the left side, select all injections in the sequence. 
 
 8.3.7 Under Methods on the left side, right-click on the processing method and select
 
 “Link selected injections to selected method”. 
 
 8.3.8 At the top right, select “Injection List” if it is not already selected.
 
 8.3.9 The Injection List will be displayed in the main window. 
 
 8.3.10 At the top of the Injection List, change the bracketing mode to Overall.
 
 8.3.11 Change Sample Type to “Cal. Std.” for all working standard injections.
 
 8.3.12 For each working standard injection, select the “Compound Amounts” column
 and enter the standard concentration in g/mL under “Amount in Sample”. The
 
 standard concentration calculation should be included in the laboratory notebook.
 
 Do not round or truncate the calculated value. 
 
 8.3.13 For each of the samples, enter the sample weight in g in the Sample Amount
 column. 
 
 8.3.14 For each of the samples, enter the sample volume in mL in the Dil. Factor 1
 
 column. If multiple dilutions were performed, the dilution factor should account
 for all dilutions. For example, the sample was initially dissolved in 100 mL to
 
 prepare a stock solution. Then, a 2.0 mL aliquot of the stock solution was diluted
 
 to 100 mL. In this case, the dilution volume is: 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | D-808 2 | Page 14 of 18 
 
 and Reporting 
 
 00 mL 
 Dilution Volume = 100 mL x = 5,000 mL 
 2mL 
 8.3.15 For each of the samples, enter the weight of a single dosage unit in g in the Unit
 
 Dose column. 
 
 8.3.15.1 For capsules: take the average capsule weight from the weight variation
 test and subtract the weight of the capsule shell listed in the product
 
 profile. Alternatively, combine the fill material from at least 10 capsules,
 record the weight of the combined fill material, and calculate the average
 
 fill weight. 
 
 8.3.15.2 For tablets and gummies: record the weight of at least 10 tablets, and
 
 calculate the average tablet weight. 
 
 8.3.15.3 For liquids, powders, and capsule-in-capsule products: use the unit dose
 weight listed in the product profile. 
 
 8.3.15.4 For raw materials, enter 1. 
 
 8.3.16 For each of the samples, click on the Compounds Custom Fields column, and
 
 enter the label claim amount in mg for each target analyte in the Label Claim
 column. For raw materials, enter 1. 
 
 8.3.17 For each of the samples, enter any correction factors needed in the Multiplier 1
 
 column. For example, a multiplier is required if the reference standard is a
 
 hydrochloride salt and the sample must be reported as the tartrate salt. Common
 conversion factors are listed in D-903. 
 
 8.3.18 At the top, click the Reprocess All icon ®. 
 
 8.3.19 If the chromatograms window is not open, click on Chromatograms at the top.
 
 8.3.20 Click through each standard or sample on the left side and examine the
 
 chromatogram to ensure that it is integrated properly. 
 
 8.3.21 If any chromatogram is not integrated properly: 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | -808 2 | Page 15 of 18 
 
 and Reporting 
 
 8.3.21.1 Ifthe processing method window is not open, click Processing Method
 
 at the top. 
 
 8.3.21.2 Adjust the integration events, and reprocess data until all
 chromatograms are properly integrated. 
 
 8.3.21.2.1 Adjustment of the Slope Sensitivity and Peak Width is often
 
 sufficient to resolve integration issues. For more advanced
 integration techniques, consult the manual by choosing
 
 Agilent Technologies > OpenLab Help & Learning from
 the Windows Start Menu. DO NOT USE MANUAL 
 
 INTEGRATION. 
 
 8.3.22 If all chromatograms have been integrated properly, click Save All Results.
 
 8.4 Creating a Quantitative/Qualitative Results Report 
 
 8.4.1 After all integrations are satisfactory and results are saved, click on Reporting at
 
 the lower left corner. 
 
 8.4.2. On the left side: select the standards/samples to be included in the report, and
 double click the desired report template. Report templates are designed for either
 
 finished product or raw materials (designated by FP or RM). 
 
 8.4.3 Filtering Reported Results (if required) 
 
 8.4.3.1 At the top, click Editor. 
 
 8.4.3.2 Right click the table to be edited, and select Properties.
 
 8.4.3.3 On the left side, select Filtering. 
 
 8.4.3.4 To filter by sample type: 
 
 8.4.3.4.1 Expression = Sample_Type 
 
 8.4.3.4.2 Operation = (or <> for not equals) 
 
 8.4.3.4.3 Value = 1 for Cal. Std. or 3 for Sample 
 
 8.4.3.5 To filter by sample name or compound name: 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Use of OpenLab for HPLC and GC Data Acquisition D-808 
 Page 16 of 18 
 and Reporting 
 
 8.4.3.5.1 | Expression = Sample_Name (or Compound_Name)
 
 8.4.3.5.2 Operation = (or <> for not equals or contains)
 
 8.4.3.5.3 Value = “*STD*” (example for any sample with STD in the
 
 name, * is a wildcard) 
 
 8.4.3.6 Results can be filtered on multiple parameters. For more information on
 filtering techniques, consult the manual by choosing Agilent
 
 Technologies — OpenLab Help & Learning from the Windows Start
 Menu. 
 
 8.4.3.7 Click on the Save Template icon at the top of the page.
 8.4.4 Click on the Preview Icon re at the top of the page. 
 
 8.4.5 Click on the Refresh Preview Icon Ss at the top of the page.
 
 8.4.6 Ifthe report preview is acceptable, print the report by selecting the Print Icon a
 
 at the top of the preview window. 
 
 8.5 Creating an Injection Report 
 
 8.5.1 On the left side: select all injections in the sequence. 
 
 8.5.2 Double click the appropriate injection report which corresponds to the instrument
 that the sequence was run on. 
 
 8.5.3. Click on the Preview Icon AG at the top of the page. 
 
 8.5.4 Click on the Refresh Preview Icon at the top of the page. 
 
 8.5.5 Print the report by selecting the Print Icon © at the top of the preview window.
 
 8.6 Creating a Chromatogram Report 
 
 8.6.1 On the left side: select a blank injection, a standard injection, and the sample
 
 injection. 
 
 8.6.2 Double click the Chromatogram report. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Use of OpenLab for HPLC and GC Data Acquisition | -808 2 | Page 17 of 18 
 
 and Reporting 
 
 8.6.3. Click on the Preview Icon Ml at the top of the page. 
 
 8.6.4 Click on the Refresh Preview Icon Cy at the top of the page.
 
 8.6.5 It may be necessary to change the filtering of the chromatogram to display the
 
 chromatogram for the appropriate wavelength. 
 
 8.6.5.1 Click Report Editor at the top of the screen. 
 
 8.6.5.2 Right click on the chromatogram, and select Properties.
 
 8.6.5.3 Select Filtering. 
 
 8.6.5.4 Under Expression, choose Signal_Description. 
 
 8.6.5.5 Under Operation, select Contains. 
 
 8.6.5.6 Under Value, enter a unique part of the signal description surrounded by
 
 asterisks (e.g. *225* or *DAD1A*). 
 
 8.6.5.7 Click Refresh Preview. 
 
 8.6.6 It may be necessary to change the scaling of the chromatogram so that the peaks
 are more clearly visible. 
 
 8.6.6.1 Click Report Editor at the top of the screen. 
 
 8.6.6.2 Right click on the chromatogram, and select Properties.
 
 8.6.6.3 Select Signal Axis. 
 
 8.6.6.4 Select Scaling > All signals in given scale. 
 
 8.6.6.5 Enter the desired minimum and maximum range of the y-axis.
 8.6.7 Print the report by selecting the Print Icon © at the top of the preview window.
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Use of OpenLab for HPLC and GC Data Acquisition D-808 2 | Page 18 of 18 
 
 and Reporting 
 Revision Date Description of Changes CCR# By 
 901 1/Bevision History N/A S. Sassman
 1 04H4/2; S-Sassman
 
 0 New 
 01/16/25 Update to reflect current lab practices. CC-23-0181 M.| Maples
 Add Section 6 General Documentation and Reporting Guidelines and
 form D-808-F1 
 2 Update Section 7.10 and 7.10.8 for clarity Ce2m2 
 Update Section 7.11 to reflect correct method 
 
 
 

[SOP 

[No text on page 19]