1.0 Purpose 
 
 The purpose of this procedure is to define the process for the cleaning and operation of the
 autoclaves located in the QC Laboratory. 
 
 2.0 Scope 
 
 The procedure covers the weekly and monthly cleaning and quarterly testing of the autoclaves
 when in use. Specific procedures are also given for sterilizing wrapped and unwrapped
 
 utensils, glassware, plastics, and liquids. 
 
 3.0 Responsibility 
 
 301 It is the responsibility of QC Laboratory personnel to follow this procedure.
 
 3.2 It is the responsibility of QC Laboratory Management to implement this procedure and
 to ensure that the procedure is being followed. 
 
 3.3. It is the responsibility of QC Laboratory Management or designee to keep this
 procedure aligned with current practices. 
 
 4.0 Definitions 
 
 4.1 QC — Quality Control 
 
 4.2. PQV-— Process Quality Verification 
 
 5.0 References 
 
 5.1 Tuttnauer EZ10 Operation and Maintenance Manual 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Operation and Cleaning of the D-824 Page 2 of 7 
 Autoclaves in the QC Laboratory 
 
 5.2. Tuttnauer 3850EA Operation and Maintenance Manual 
 
 5.3 C-502, SOP, Record Storage, Retention, and Destruction 
 
 5.4 C-501, SOP, Document Control Procedure 
 
 53 A-106, SOP, Documentation Guidelines for cGMP Records 
 
 5.6 D-824-F1, Form, Autoclave Usage Log 
 
 5.7 D-824-F2, Form, Autoclave Cleaning Log 
 
 5.8 D-824-F3, Quarterly Testing Log for Autoclave 
 
 6.0 Procedure 
 
 6.1 Weekly Cleaning 
 
 6.1.1 Each week of operation, the below defined tasks must be performed. Document
 
 on Form D-824-F2 Autoclave Cleaning Log. 
 
 6.1.1.1 Cleaning the chamber and trays 
 
 6.1.1.1.1 Drain the water from the reservoir using the drain tube
 located on the front of the unit. 
 
 6.1.1.1.2 Wash the inside of the chamber and trays with mild soap or
 
 Speed-Clean and Millipore water. 
 
 6.1.1.1.3 Refill the reservoir with Millipore water. 
 
 Note: Do not use abrasive or bleaching agents in the chamber,
 such as scouring powder, steel wool, bleach, or wire
 
 brushes. 
 
 6.2. Monthly Cleaning 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 Operation and Cleaning of the D-824 8 Page 3 of 7 
 Autoclaves in the QC Laboratory 
 
 6.2.1 Each month of operation, the below defined tasks must be performed.
 
 Document on Form D-824-F2 Autoclave Cleaning Log. 
 
 6.2.1.1 Cleaning the chamber and plumbing 
 
 6.2.1.1.1 With a cooled chamber, drain the reservoir and fill it with
 
 clean Millipore water. 
 
 6.2.1.1.2 Add one packet of Chamber Brite autoclave cleaner directly
 
 to the bottom of the chamber. 
 
 6.2.1.1.3 Run one “Wrapped” cycle with dry time set at zero.
 
 6.2.1.1.4 Drain the reservoir and refill it with Millipore water.
 
 6.2.1.1.5 Rinse by running one “Unwrapped” cycle. 
 
 6.2.1.1.6 After the cycle has completed, drain and refill the reservoir
 with Millipore water. 
 
 6.2.1.1.7 Allow the sterilizer to cool. 
 
 6.2.1.1.8 Remove the trays and tray rack and wipe off with a damp
 
 cloth. 
 
 6.2.1.1.9 Wipe out the inside of the chamber, using care not to
 
 damage the heating element, steam temperature probe, or
 
 level sensor probe. 
 
 6.2.1.1.10 Reinstall the tray rack. 
 
 6.3 Quarterly testing for Bacillus stearothermophilus ATCC 7953 to ensure steam
 production 
 
 6.3.1 Each quarter of operation, the below test is required to be performed. Document
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 
 Operation and Cleaning of the D-824 8 | Page 4 of 7 
 Autoclaves in the QC Laboratory 
 
 on Form D-824-F3 Quarterly Testing Log for Autoclave. 
 
 6.3.1.1 Place a test strip in autoclave (10° spore population) in a small sterile
 tube with cap. Fill with TSB broth, making sure to completely
 
 submerge the strip. 
 
 6.3.1.2 Press the program “3” button. 
 
 6.3.1.3 Press the “Start” button. 
 
 6.3.1.3.1 Conditions: 250°F (121°C), 16psi, sterilize 25min; dry
 25min, slow vent. 
 
 6.3.1.4 Allow all liquids to cool for one hour before handling.
 
 6.3.1.5 Prepare a second sterile tube with a test strip and TSB broth, but do not
 
 autoclave. Place both tubes in a heating block set at 55°C for 5 days.
 
 6.3.1.6 Analyze the autoclaved test strip for the presence of Bacillus
 
 stearothermophilus. The test strip that was not autoclaved should
 display luxuriant growth; if there is no growth, assess whether there
 
 was a problem during sample preparation or incubation and repeat
 
 testing. 
 
 6.4 Sterilizing unwrapped instruments (Program 1) 
 
6.4.1 When appropriate, attach a piece of autoclave tape to each instrument.

 6.4.2 Place the instruments or tubing loose on the tray. 
 
 6.4.3 Press the “Unwrapped” button. 
 
 6.4.4 Press the “Start” button. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 
 Operation and Cleaning of the D-824 8 | Page 5 of 7 
 Autoclaves in the QC Laboratory 
 
 6.4.4.1 Conditions: 270°F (132°C), 28psi, sterilize 3min; dry Omin.
 
 6.5 Sterilizing wrapped instruments and porous loads (Program 2) 
 
 6.5.1 Place the instruments or tubing in an appropriate sized autoclave pouch with an
 
 indicator and seal. 
 
 6.5.2 Place the pouched or loosely wrapped instruments on the tray.
 
 6.5.3. Press the “Wrapped” button. 
 
 6.5.4 Press the “Start” button. 
 
 6.5.4.1 Conditions: 270°F (132°C), 28psi, sterilize 7min; dry 30min.
 
 6.6 Sterilizing glassware (Program 3) 
 
 6.6.1 Wrap the instruments in aluminum foil, cover the openings of the
 glassware/plasticware with aluminum foil, cover capped bottles with aluminum
 
 foil and make sure the cap is loosened % turn. 
 
 6.6.2 Attach a 1 to 2 inch strip of autoclave tape to each piece on the aluminum foil.
 
 6.6.3 Place the packs, glassware and/or plasticware loosely spread out on the tray.
 
 6.6.4 Press the “Glassware” button. 
 
 6.6.5 Press the “Start” button. 
 
 6.6.5.1 Conditions: 250°F (121°C), 16psi, sterilize 30min; dry Omin.
 
 6.6.6 Allow all glassware to cool for fifteen minutes before handling.
 
 6.7 Sterilizing liquids 
 
 6.7.1 Cover all openings on glassware containing liquids with aluminum foil, cover
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 
 Operation and Cleaning of the D-824 8 | Page 6 of7 
 Autoclaves in the QC Laboratory 
 
 capped bottles with aluminum foil and make sure the cap is loosened % turn.
 
 6.7.2 Place containers with liquids loosely on tray. Do not exceed 2.0 liter of total
 liquid volume to be sterilized.in the small autoclave, and 4.0 liter of total liquid
 
 volume to be sterilized in the large autoclave. Place autoclave tape on outside
 
 of container. 
 
 6.7.3 Press the program “3” button. 
 
 6.7.4 Press the “Start” button. 
 
 6.7.4.1 Conditions: 250°F (121°C), 16psi, sterilize 25min; dry 25min, slow
 
 vent. (PSI and Time may vary according to liquid being processed, see
 manufacturer instructions) 
 
 6.7.5 Allow liquids to cool for one hour before handling. Check to see that autoclave
 tape indicates sterilization. 
 
 6.8 Logging use and cleaning of the Tuttnauer Autoclaves 
 
 6.8.1 Prior to running the autoclave, complete the log book entries on Form D-824-F1
 Autoclave Usage Log for description of process, program, temperature,
 
 pressure, time, analyst and date. 
 
 6.8.2 At the completion of the autoclave cycle, review the tape and check that during
 
 the sterilization phase the temperature did not exceed +/- 4°C from target and the
 pressure did not exceed +/- 2psi the operation temperature and pressure. Log
 
 information on Form D-842-F1 Autoclave Usage Log. 
 
 6.8.2.1 Alternatively, if no print-out is available from the autoclave, ensuring
 
 the autoclave tape indicates sterilization was completed is also
 
 acceptable. 
 
6.9 Refer to manufacturer’s operation and maintenance manual for complete instructions on

 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Operation and Cleaning of the D-824 Page 7 of 7 
 Autoclaves in the QC Laboratory 
 
 programming, cleaning and operating the autoclave. 
 
 6.10 Documentation will be maintained following SOP C-502 Record Storage, Retention,
 
 and Destruction and C-501 Document Control Procedure. 
 
 6.11 A PQV check must be performed for each completed logbook page as outlined in SOP
 
 A-106 Documentation Guidelines for cGMP Records. 
 
 7.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 1 | 01/17/13 | New - - | - | - |
| 2 | 04/03/13 | Changed SOP log. Added section 5.9. Changed monthly testing to quarterly testing in section 5.3 12/02/14 Updated format. Added Form D-824-F2. Updated Form D-824- 14-0969 B. Johns F1. Updated responsibilities. | 13-238 | B. Johns |
| 4 | 04/04/17 | Biennial review: Changed conditions in steps 6.3.3.1 and 6.7.4.1 Updated for new autoclave model and changed title to reflect. p) 04/18/18 Clarified quarterly testing for Bacillus stearothermophilus ATCC 18-0128 E. Johnson 7953. | 16-1101 | N. Zhan_g |
| 6 | 06/01/21 | Updated SOP to include both models of Autoclave CC- Updated SOP to ensure time is included as well as PSI, Added that | 21-0225 | G. Shaw |
| 7 | 02/02/23 | PSI and time can be changed according to manufacturer CC- instructions on liquid being autoclaved. Added Form D-824-F3 for clarity on quarterly testing. Other | 23-0060 | G. Shaw |
| 8 | 10/02/23 | minor edits. Made SOP more general. Added documentation CC- requirements. Updated logo. | 23-0514 | J. Sassman |

 
 

[SOP 

 Autoclave Usage Log 
 ION 
 py ruztiona Form: D-824-F1 CCRNo. -CC-23-0514 Revision: 8 
 
 Logbook Number: Logbook Page: 
 
 Pass Ol Fail 
 
 Items Autoclaved: 
 Item Name Lot Number Expiration Date Qty Size Notebook #/Page# 
 
 Comments: 
 
 Performed By/Date: 
 
 Reviewed By/Date: 
 
 