F-503
Millipore Water System
Original Document
Scanned document (image-only PDF)
Extracted Text
Searchable text extracted from PDF
1.00 Purpose The purpose of this procedure is to provide instruction on the operation and maintenance oftthe Integral 5 Milllgarne Water System. 200 Scope This procedure applies to the Milligore Milli-Q Integral Water Pumification System for use in the QG Laboratory at lon Labs, Inc. 300 Regpomsibility 811 itis the reqnonsitility of @C Laboratory Personnel to parfonm daily, or as needed inspections of the system, to change the cartridges during scheduled replacement, and to perform periodic system flushes. 222 Itis the respomsitility of @G Laboratory Management to implement this procedure and to ensure that the procedure is being followed. 833 It is the responsibility of @G Laboratory Management to keep this SOP current with latest lon Lalbs practices. 400 Definitions 411 QE = Quality Control 500 Referennes 511 Milligore User Manual, Milli-Q Integral 3/5/10/15 systems &22 §E-705, SOP, Water Use, Sampling, and Testing F3i3 EOSE2, Fonm, Purified add Millipore Water Test Ticket = Microbial Limit gHComathotwitiOtOg 7C 514 §£-705-3, Form, Conductivity Stage 2 Test Ticket 600 Procedure try1 are me [SOP Standard Operating Procedure SOP No Rev Milllipone Water System F-503 6 Page 4 of 4 6BELA4 Select “Memu" then “Samitize/Clear" amd them select “RO C12 Cleaning", the process will last anpronimatbly 19 6.3.6.3 RO pHi Cleaning = anproximatdly evesy 6 months im place of CL2 Cleaning, 6363311 Open the Sanitization port and place a ROClean pouch into the port. 636322 Put the cap back on and tighten it. 636333 Go into “Standby” mode 6363314 Select “Menu” then select “Sanitize/Clean” 636335 Select “RO pH cleaning”, the cleaning will last 742 minutes. 6363H6 Remove the ROClean pouch the next time the Sanitization port is opened Ne@te: Do not petfom CL2 and pH cleanings at the same time. fill Cleaning is used as a supplement to the CL2 cleaning 644 Step-By-Step MRintemamceinsttuctionss 6411 Refer to the Milliq@ore User Manual, Milli-Q Integral 35/10/15 systems for step- by-step instructions on replacing cartridges and performing system maintenance checks and flustiess. 700 Revision History Revision Date Descrition ofChany CCR # By 0 07/01/10 Original - - | OvEDtO Removed conductivity testiDg requirementts, changed testing _ ; fiom each tank change to quarterly checks O4/OIS | r U e p f d l a ec t t e t d h e S O Mi P l l f i o g r o m n a e t , w a n t a e m r e , s y a s o t d e m S OP number, changed SOP to = ~ B. Johos Biennial review: Updated! SOP fornratt. Added consumabie part 3 ON/05/16 #s. Expanded information for routine inaimtemansse and dail) 16-0024 N. Zhang inspections. 4 @4i08119 | Scheduled review: added items to checklist. 19-O239 J. Maiigman Revise SOP to reflect current practioes and sdded sanitization ; . 2/23/19 steps per instrument manual 19-0964 I. Garett Included Reference to SoP for Mille Water Testing and oOROD 6 oR26/20 removed speuifications ofsystem that are not needed — L. McWade CONFIDENTIAL; For ION Labs use only [SOP Standard Operating, Procedure SOPNo | Rev Millipore Water System F-503 6 Page 3 of 4 62229 ROCare A Pouches = Part# ZW AQIDOI2 633 Routinemainttanance 6311 The maintenamoe schedutte i§ autonmateti for the Millipore Integral 5 Water purification system. Alerts willl appear on the display when a cartridge change, lamp change or cleaning process is nearing the maintenance date. The indicator light will tum yellow when an alert is active. 63222 When the system is out of spavification and/or maintenance is due the indicator light vill tum red! amd! water prodwation willl be suspended umiil the alert is addressed. 63333 The system will need to be put into stamdby mode when a cartridge is changed or a clearihg procedure is being followed. The alert will need to be reset at the completion of the task before water production rvill resume. 63344 Document ail maintenance ant sewice, includimg cartridge chamges im the Milligane Water logbook. 62335 Consumable Replacement Schedule: Consumable Replacement Intervals Proguard Pak aint oe Every 6 months or when the system check 22 pm POD Filter recommends a r e e p pl l a a c c e ement Tank Vent Filter EDM Module Checked or r eplaced when alarm dictates th that cunent is not maintained at needed level UV Lamps Every 3 years or when the system check recommends a replacement 63%6 Sanitization of RO Cartridge 6.3.6.1 Perform RO Cartridge samitization in 3 and 6 month intervals 6.3.6.2 RO CL2 Cleaning = approximately every 90 days 6862211 Open Samitizatiom Port and place a chlorine tablet in the poft. 686222 Put the cap back on and tighten it 63223 Go into “Standby” Mode [SOP Standard Operating Procedure ve No Rev F-503 _Mijllipore Water System 6 Page 2 of 4 “4411p; 61.1 Daily Inspections 6.i.ll Ready mode as indicated by a grean light should! be displayed om the Main Unit’s LED sereen when normal water production and delivery are required. 61122 A yellaiw light will display as a caution to notify the user of impending serwice requirements such as a cartridge change or system flush. 61133 After turning on water flow at the @-BOB gun and check the indicator display on the @-POp during dispensing to ensure that no warnings are displayed and water metrics are within tolerance. 6.113311 The reservoir is represented by ten graphic bars, each bar represemting 10% capacity ofpuified water in the tank. 6.11322 Resistivity should read > 18MQ. 6113233 TOC oxidizable carbon level <10 ppb. 6.113344 Water temperature should be near ambient. 6.113335 The Proguard Pak is in use as indicated by the symbol. 6.1.866 The Quantum Cartridge is in use as indicated by the symbol. 6.113377 No wamings or maintenance alerts are displayed. 6.14 Inspect the system for leaks. 22 Consumabies 62211 Proguard Pak = Pafi# PROGOTOS2US 62222 Bactericidal UV Lamp- Pant# ZLXUVYILIR} 62233 Photooxidation UV Lamp- Part# ZMQUVILPO1 62244 Quantum Cartricige - Part QTUMOTEX!! 6225 AlO TOC Monitor Lamp- Part# ZFAIOUVOI 6246 POD2Z24iM Filter- Part# MPOPOAGO 6277 Tank Filter- Part¥ TANKMPKOI 62248 Samitization Tablets (US)- Part# 5874316024 CONFIDENTIAL; For ION Labs use only