G-201

Calibration Program

Section G — Equipment Revision 6 14 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 The purpose of this procedure is to describe the process for the identification, scheduling, and
 logging of equipment to be calibrated. 
 
 2.0 Scope 
 
 This procedure applies to all equipment maintained by Ion Labs, Inc. requiring calibration. This
 procedure describes general calibration requirements and is superseded by instrument/machine
 
 calibration requirements described in its specific procedure. Furthermore, analytical instrument
 calibrations are governed by SOP G-208 Analytical Calibration and Equipment Control and
 
 any calibration requirements described in that procedure supersede these instructions.
 
 3.0 Responsibility 
 
 cm It is the responsibility of Analytical Development Management to align this procedure
 
 with current practices and to delegate and oversee management of the program.
 
 3.2 It is the responsibility of Maintenance to submit operational equipment profiles to
 
 document control, and the Metrologist, or designee, for new analytical equipment as
 outlined in SOP G-208, Analytical Calibration and Equipment Control.
 
 Due It is the responsibility of Quality to oversee and ensure compliance to this procedure for
 
 all Quality Personnel. 
 
 3.4 It is the responsibility of Production to oversee and ensure compliance to this procedure
 
 for all Production Personnel. 
 
 3.5 It is the responsibility of the QC Laboratory and Metrologist to oversee and ensure
 
 compliance to this procedure for all QC Laboratory Personnel. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev Baseman 
 Calibration Program G-201 6 age 20 
 
 3.6 Any qualified individual may arrange or execute calibrations, place calibration stickers
 on equipment, however, logging and maintaining equipment calibration schedules is
 
 performed by the Metrologist or designee. 
 
 4.0 Definitions 
 
 4.1 Calibration — Comparison of equipment performance against a traceable standard or
 
 certified instrument to detect, correlate, report or eliminate by adjustment, any variation
 in the accuracy of the item being compared 
 
 4.2 Certification — Written verification that the device operates within applicable
 specifications 
 
 4.3 Verification — An equipment check against a standard to confirm that the equipment is
 in a state of calibration (or operating as designed where applicable). The equipment can
 
 remain in operation if the calibration is verified as within specification for a period not
 
 to exceed the recommended verification interval. 
 
 4.4 Due Date — The month the piece of equipment needs to be calibrated or verified.
 
 Equipment due in a specific month needs to be calibrated or verified by the last day of
 the month of the next calibration due date. The instrument may be used within that
 
 month. The equipment is not to be used on or after the first day of the next month
 without a certificate of calibration in effect. In extreme circumstances, an item is used
 
 outside its calibration date. In that case, the results gained from the equipment may be
 
 deemed valid if the item upon next calibration has its “as found” results within
 specifications. 
 
 4.5 QC — Quality Control 
 
 4.6 QA — Quality Assurance 
 
 4.7 AD — Analytical Development 
 
 4.8 PM — Preventative Maintenance 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev a EaCRNOEAT 
 Calibration Program G-201 6 age 3 0 
 
 4.9 OQ — Operational Qualification 
 
 4.10 PQ — Performance Qualification 
 
 4.11 Asset Essentials — a cloud-based software that enables you to assign and track the
 progress of all maintenance activities, manage equipment uptime and reliability. For
 
 calibration purposes, Asset Essentials is utilized for scheduling only.
 
 5.0 References 
 
 dul G-102-F1, Form, Equipment Profile 
 
 Die G-201-F1, Form, Calibration Certificate 
 
 5:3 G-201-F2, Form, Completed Calibration Log 
 
 5.4 QS-114, SOP, Risk Management and Risk Analysis 
 
 3.5 G-208, SOP, Analytical Calibration and Equipment Control 
 
 5.6 C-502, SOP, Record Storage, Retention, and Destruction 
 
 5.7 ANSI/NCSL Z540.3-2006; ISO 10012-1, ISO/IEC Guide 25 
 
 6.0 Procedure 
 
 6.1 Asset Essentials: 
 
 6.1.1 New assets that require calibration are logged into Asset Essentials, typically
 
 initiated by form G-102-F1 Equipment Profile. 
 
 6.1.2 Information entered into Asset Essentials: 
 
 6.1.2.1 Description of Equipment 
 
 6.1.2.2 Ion Asset Number 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 G-201 6 Page 4 of 11 
 Calibration Program 
 6.1.2.3. Location of equipment (if needed) 
 
 6.1.2.4 Calibration Due Date 
 
 6.1.3 Current Assets already cataloged in Asset Essentials are amended as calibrations
 are completed and new calibration due dates are assigned. 
 
 6.1.4 Assets are typically cataloged/ grouped by calibration due date month in Asset
 Essentials. 
 
 Monthly, Asset Essentials will generate a work order for all required
 calibrations due in that month. This work order is emailed to the metrologist or
 
 designee for completion. 
 
 6.1.6 Upon successful completion of calibrations, the work order can either be closed
 
 or amended with a new due date in Asset Essentials. 
 
 6.1.7 Form G-201-F2 Equipment Calibration Log will be completed as needed to log
 calibration information for equipment. This form will be cataloged by month/
 
 year and maintained in a binder or equivalent storage format. The form will
 include the following information: 
 
 6.1.7.1 Month 
 
 6.1.7.2 Year 
 
 6.1.7.3 Ion Asset # 
 
 6.1.7.4 Description 
 
 6.1.7.5 Calibration type - Temperature, Humidity, Weight, RPM, etc.
 
 6.1.7.6 Frequency - Weekly, Monthly, Biannual, Annual, etc. 
 
 6.1.7.7 Internal or External calibration 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 G-201 6 Page 5 of 11 
 Calibration Program 
 
 6.1.7.8 Date of calibration 
 
 6.1.7.9 Logged by/Date 
 
 6.1.7.10 Page 
 
 6.1.8 Each sheet of form G-201-F2 Equipment Calibration Log can correspond only
 
 to one month and one year. At the end of each month, the unused part of the
 form will be lined out, initialed and dated. 
 
 6.1.9 Stored within each monthly section is a calibration certificate for every piece of
 equipment tested during that specific month. 
 
 6.1.9.1 External calibration certificates will be labeled with the Ion# and/ or
 
 equipment serial number at the top of each page for tracking purposes,
 if not already present on certificate. 
 
 6.1.9.2 The calibration due date on the external calibration certificate should
 reflect at least the minimum frequency specified by the Ion Labs
 
 calibration/verification program otherwise the Ion Labs frequency
 supersedes the vendor frequency. Any changes made to the calibration
 
 due date on the certificate must be signed or initialed.
 
 6.1.10 Equipment that either does not pass or has an expired calibration can either be
 removed from service and obsoleted or sent for repair. If repaired, a new
 
 calibration must be completed prior to placing back into service.
 
 6.1.11 For new equipment accompanied by a calibration certificate and calibrated by
 
 the manufacturer, the manufacturer due date will supersede Ion Labs Frequency
 specifications. For subsequent calibration verifications, the Ion Labs frequency
 
 will be used. 
 
 6.1.12 Certificates for scales, weights, thermometers, and calipers are retained by
 
 metrology and are also referenced in all equipment profiles in Asset Essentials.
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 G-201 6 Page 6 of 11 
 Calibration Program 
 6.2 Execution of the Calibration Program 
 
 6.2.1 The program will be overseen by the Metrologist or designee.
 
 6.2.2 Form G-201-F2 Equipment Calibration Log will be filled out and closed
 monthly. 
 
 6.2.3 PM work orders for analytical equipment in production will be generated by
 Asset Essentials and performed by the metrologist or designee. Analytical
 
 equipment in production includes: Friabilator and hardness instruments used for
 in-process product evaluation. 
 
 6.2.4 Purchase orders will be prepared from quotes for equipment to be tested by an
 
 external calibration service. 
 
 6.2.5 All calibration vendors must be approved by QA or Metrology before use.
 
 6.2.6 For equipment calibrated by external sources, a copy of the calibration
 certificate will be provided by the service provider and forwarded the
 
 metrologist or designee for filing. 
 
 6.2.7 If applicable, Calibration/PM performed by external vendors must meet any
 
 requirements for calibration stated in the equipment’s profile and/or SOP.
 
 6.2.8 A qualified individual, preferable the metrologist, will review all external
 
 calibration certificates and write Ion# and/ or serial number at the top of the
 
 certificate, if not already present. 
 
 6.2.9 All certificates received externally will contain a decision as to the condition “‘as
 
 left”, completed by provider. This decision may include a “Pass/ Fail”
 statement or an “In Tolerance” statement. 
 
 6.2.9.1 External service providers used for calibration activities must provide
 this decision statement. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev ee 
 Calibration Program G-201 6 age 7 0 
 
 6.2.10 For in-house calibrations, Form G-201-F1 Calibration Certificate with the Ion#
 
 and description listed will be issued to the person who completes the calibration.
 
 6.2.11 The designated person will record all appropriate information about the
 calibration on Form G-201-F1 Calibration Certificate and the form along with
 
 any corresponding paperwork will be returned to the metrologist or designee.
 
 6.2.12 Completed internal forms will be reviewed by qualified personnel for final
 
 approval. 
 
 6.3 Establishing the Frequency of Calibration 
 
 6.3.1 The calibration frequencies listed in Attachment 1 Equipment Calibration
 
 Frequency Table, are minimum Ion Labs frequency recommendations.
 Deviations from the table for specific pieces of equipment may be justified by
 
 review of manufacturer's documentation and/or previous certifications.
 
 6.3.2 Early calibrations are permissible. The calibration schedule will reset beginning
 
 from the latest calibration date. 
 
 6.3.3. Certain equipment (Dickson One Temperature/ Humidity Sensors, Digital
 
 Temperature and Humidity Monitor, Conductivity meter, Jumbo Timer etc.) will
 be replaced after factory certification is expired. The integrity of the metal
 
 detector testing balls, rods and sticks will be checked every year to ensure no
 
 rusting and chipping of the metal wrapped inside. If they are without defect,
 they may be used indefinitely. 
 
 6.4 Calibration Stickers 
 
 6.4.1 Each piece of calibrated equipment will be marked with a sticker indicating its
 
 calibration status. 
 
 6.4.2 Internally placed stickers will indicate at minimum: 
 
 6.4.2.1 ION ID# 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 G-201 6 Page 8 of 11 
 Calibration Program 
 6.4.2.2 The initials of the person conducting the calibration, or applying
 
 the sticker. 
 
 6.4.2.3 The date of the calibration (MM-DD-YY) 
 
 6.4.2.4 The due date for the next calibration (MM-YY) 
 
 6.4.3. Externally completed calibrations will still be tagged using internal Ion
 calibration stickers for consistency. 
 
 6.4.3.1 External company stickers may also be present on calibrated
 equipment in conjunction with Ion stickers. 
 
 6.5 Failed Calibration 
 
 6.5.1 A piece of equipment that does not meet its calibration requirements will be
 
 removed from service and logged as per Section 6.1.8. 
 
 6.5.2 If the equipment is serviceable an external calibration/repair service will repair
 and recertify the piece of equipment before returning to service.
 
 6.5.3. A"Do Not Use" placard will be affixed to all pieces of equipment that cannot be
 internally verified and the sticker will only be removed after it has been repaired
 
 and recertified by an approved vendor. 
 
 6.6 Failed as Found Calibration 
 
 6.6.1 If equipment does not meet its as found specification an investigation/risk
 assessment should be performed as per QS-114 Risk Management and Risk
 
 Analysis to determine its effect. 
 
 6.6.2 Any instrument specific procedure instructions supersede this guidance.
 
 6.6.3 Any existing processes can be used for the investigation/risk assessment as well.
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 G-201 Page 9 of 11 
 Calibration Program 
 
 6.6.4 Ifthe effect is minimal, the calibration certificate may be used to document any
 risks. 
 
 6.7 Calibration Standards 
 
 6.7.1 Reference standards used for internal calibrations must be calibrated or verified
 
 in compliance with ANSI/NCSL Z540.3-2006; ISO 10012-1 and/or ISO/IEC
 Guide 25 using NIST traceable standards or standards derived from accepted
 
 values of naturally occurring physical constants by an external service or source.
 
 The calibration period of these items are set by the service and may not be
 extended beyond the period indicated on their certificate. 
 
 6.7.2 Standards used for any equipment calibration will be captured as part of the
 calibration paperwork. 
 
 6.7.2.1 Applicable C of A’s or certificates of calibration for standards will be
 maintained with each piece of equipment for each calibration.
 
 6.8 Document Retention 
 
 6.8.1 All documents will be maintained following C-502 Record Storage, Retention,
 
 and Destruction. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 G-201 6 | Page 10 of 11 
 Calibration Program 
 
 7.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 05/17/10 | New : - | - | - |
| 1 | 04/01/13 | Changed the SOP format & number, changed forms number : - Restructured Calibration Program. Realigned responsibilities. | - | - |
| 2 | 11/05/13 | . Improved clarity of the process. Added a Master Equipment List (electronic). Reorganized forms G-201-F1-F2-F3. Updated SOP format. Realigned responsibilities. Consolidated forms G-201-F1, G-201-F3. Added electronic spreadsheet, 3 Sh ORNS Calibration Schedule/Equipment List. Added document retention te-O022 Nichane time. Adjusted calibration frequencies for scales and weights. | 13-1013 | B. Johns |
| 4 | 04/04/17 | Add metal detector test kits and thermal and humidity monitor to program | 16-1120 | N. Zhang |
| 5 | 11/10/21 | Updated to reflect current processes. As found out of specification CC- report requires actions, Changed form numbers. Update to add Asset Essentials as a scheduling program. Updated to | 21-0088 | J. Maignan |
| 6 | 12/08/22 | reflect a metrologist. Updated to reflect current process and CC- practices. | 22-0466 | J. Sassman |

 8.0 Attachments 
 
 8.1 Attachment 1 — Equipment Calibration Frequency Table 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev 
 G-201 6 | Page 11 of 11 
 Calibration Program 
 
 ATTACHMENT 1 — EQUIPMENT CALIBRATION FREQUENCY TABLE 
 
 Scale / Balance 1 Year External 
 
 Hardness Tester 6 Months Internal / External 
 
 Digital Viscometer 1 Year Internal / External 
 
 Friability Tester 1 Year Internal / External 
 
 Melting Point Tester 6 Months Internal / External 
 i aaa coe Deco 1 Year Internal / External 
 
 Weight Standard 2 Years / 5 Years External 
 
 Weight Standard — Daily Use 1 Year External 
 
 Torque Tester 1 Year External 
 
 Vacuum 1 Year Internal / External 
 
 Caliper 1 Year Internal / External 
 
 Metal Detector 1 Year Internal / External 
 
 Tachometer 3 Years External 
 
 Gauge Block 5 Years External 
 
 Laboratory and Production Equipment Variable Internal / External 
 
 Timer 1 Year Internal / External 
 
 Pipette 6 Months External 
 
 Disintegration Tester 1 Year Internal / External 
 
 Miscellaneous Variable Internal / External 
 
 
 

[SOP 

 Standard Operating Procedure SOP No | Rev Page Lf 11 
 Calibration Program G-201 6 age llo 
 
 ATTACHMENT 1 —- EQUIPMENT CALIBRATION FREQUENCY TABLE 
 
 Scale / Balance 1 Year External 
 
 Hardness Tester 6 Months Internal / External 
 
 Digital Viscometer 1 Year Internal / External 
 
 Friability Tester 1 Year Internal / External 
 
 Melting Point Tester 6 Months Internal / External 
 
 a aaa a inti, oak Device! 1 Year Internal / External 
 
 Weight Standard 2 Years / 5 Years External 
 
 Weight Standard — Daily Use 1 Year External 
 
 Torque Tester 1 Year External 
 
 Vacuum 1 Year Internal / External 
 
 Caliper 1 Year Internal / External 
 
 Metal Detector 1 Year Internal / External 
 
 Tachometer 3 Years External 
 
 Gauge Block 5 Years External 
 
 Laboratory and Production Equipment Variable Internal / External 
 
 Timer 1 Year Internal / External 
 
 Pipette 6 Months External 
 
 Disintegration Tester 1 Year Internal / External 
 
 Miscellaneous Variable Internal / External 
 
 
 

[SOP 

 IAN LABS Calibration Certificate 
 
 Form: G-201-F1 CCR No. CC-22-0466 Revision: 1 
 
 Standard Specification 
 
 Unit of Measure 
 Test Reading 
 
 Tolerance Allowed (Range) 
 
 Minimum Allowed Reading 
 Maximum Allowed Reading 
 
 *In Tolerance As Found (Y/N) 
 
 In Tolerance As Left (Y/N) 
 * If “As Found” tolerance is N, an explanation and risk assessment is required. Assessment may be documented below or a reference to
 another document may be listed below. 
 
 Determination (check one): O Pass O Fail 
 
 

[SOP 

d 
e 
5 
a 
 
G 
Y 
S 
A 
T 
S 
I 
O 
U 
 USyoUHueoAOwqdYie]Dnby 
, 
D 
D 
T 
T 
0 
9 
9 
N 
O 
“ 
O 
O 
Y 
D 
7 
0 
1o 
-u 
dy 
CA 
n 
s 
a 
] 
' 
Oe ’ 
q 
Us 
I 
O 
N 
4 
0 
1 
O? 
 ISNavT 
O 
N 
A 
I 
G 
A 
N 
L 
L 
V 
T