G-205

Calibration Verification of Digital Micrometers

Section G — Equipment 3 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 The purpose of this SOP is to describe the procedure for a three point calibration verification of
 
 digital micrometers. 
 
 2.0 Scope 
 
 This procedure is for the calibration verification of micrometers. This SOP does not cover the
 calibration or repair of any length measuring device. 
 
 3.0 Responsibility 
 
 3.1 It is the responsibility of QC Laboratory Personnel to follow this procedure.
 
 3.2 It is the responsibility of QC Laboratory Management to ensure the procedure is current
 and is being followed. 
 
 4.0 Definitions 
 
 41 ISO — International Organization for Standardization 
 
 4.2 EC — International Electrotechnical Commission 
 
 43 NIST — National Institute of Standards and Technology 
 
 44 ANSI - American National Standards Institute 
 
 4.5 | NCSL ~ National Conference of Standards Laboratories 
 
 4.6 QC — Quality Control 
 
 5.0 References 
 
 5.1 G-201-F2, Form, Calibration Certificate 
 
 5.2 G-201, SOP, Calibration Program 
 
 
 

[SOP G-205 | Page 2 of 3]

 Standard Operatii ng Procedure SCO2P0 N8o Re3 v Page 2 of 3 
 Calibration Verification of Digital Micrometers 
 
 6.0 Procedure 
 
 6.1 Refer to SOP G-201 Calibration Program for calibration schedule and instructions on
 the documentation of the calibration verification. 
 
 6.2 It is recommended that micrometers are verified or calibrated at an external calibration
 
 laboratory at least once every five years. 
 
 6.3 If a micrometer fails internal calibration verification and the instrument is able to be
 calibrated the unit must be calibrated and recertified by an external laboratory before
 returning to service. 
 
 6.4 Gauge block standards: 
 
 6.4.1 A gauge block set is to be used as reference standards. The gauge blocks must
 
 be verified against NIST traceable standards that have been externally calibrated
 and verified in compliance with ISO/IEC Guide 25 and/or ANSI/NCSL Z540.3-
 2006 requirements or newer. 
 
 6.4.2 Gauge blocks intended for use in calibrating micrometers require certification
 every five years. 
 
 6.5 Verification process 
 6.5.1 All measurement readings are recorded in millimeters. Change unit of measure
 
 if necessary. 
 
 6.5.2 Zero the micrometer prior to calibration verification. 
 
 6.5.3. Perform verification at three points across the range of the micrometer. Typical
 narrow range micrometer range is 1.00mm to 10.00mm and wide range
 
 micrometer range is 1.00mm to 100mm. Select gauge blocks representative of
 the low, middle and high ranges of the micrometer. 
 
 6.5.4 All three test lengths must be within tolerance to pass verification.
 
 6.5.5 All instrument information, calibration verification and environmental data
 should be documented on form G-201-F2 Calibration Certificate.
 
 6.5.6 The common tolerance for verification of narrow range micrometers is +/-
 
 0.05mm and for Wide Range Micrometers +/- 0.1mm unless otherwise specified
 by the needs of the application or recommended by the equipment manufacturer.
 
 6.5.7 The verification cycle for micrometers is 1 year. 
 
 
 

[SOP G-205 | Page 3 of 3]

 Standard Operating Procedure SOP No Rev 
 G-205 Page 3 of 3 
 Calibration Verification of Digital Micrometers 
 
 6.5.8 If a micrometer is removed from service for any reason including failed
 verification, the device in the Calibration Schedule must be lined out with the
 notation “removed from service.” The device is no longer permitted in GMP
 work areas until recertified. 
 
 6.5.9 For devices that pass verification a calibration sticker will be attached and the
 
 unit returned to service. 
 
 7.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 10/06/10 | New - - | - | - |
| 1 | 04/01/13 | Changed the SOP format & Number 2 O7/OVS caitraion toverfeaRetmoiveod fno,n 15-0879 B. Johns | 13-246 | B. Johns |
| 3 | 04/08/19 | { Scheduled review: No changes | 19-0242 | J. Maignan |