G-206

Verification of Digital Temperature and Humidity Recording Devices

Section G — Equipment Revision 3 4 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 Most digital temperature and humidity recording devices do not allow true calibration since the
 
 readings cannot be altered to match a standard device. Therefore, the readings of each device
 can be verified against a single device whose performance characteristics are certified against a
 NIST traceable standard. 
 
 2.0 Scope 
 
 This procedure can be applied to any temperature, humidity, or combination recording device.
 
 This procedure is not intended for non-recording temperature and humidity devices.
 
 3.0 Responsibility 
 
 3.1 The QC Chemists are responsible for the application of and compliance to this
 procedure. 
 
 3.2 The QC Laboratory Management is responsible for the execution of and compliance to
 this procedure. 
 
 3.3. The QC Laboratory Management is responsible for keeping the SOP current with latest
 Ton Labs practices. 
 
 4.0 Definitions 
 
 4.1 QC- Quality Control 
 
 °C — Degrees Celsius 
 
 4.2 RH-~ Relative Humidity 
 
 4.3 °F ~ Degrees Fahrenheit 
 
 5.0 References 
 
 5.1 G-201-F2, Form, Calibration Certificate 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Verification of Digital Temperature and Humidity G-206 3 Page 2 of 4 
 Recording Devices 
 
 5.2 G-201, SOP, Calibration Program 
 
 6.0 Specifications 
 
 6.1 TP125- Temperature and Humidity Data Logger, Digital 
 
 6.1.1 Operating RH conditions: 0-95% RH non-condensing 
 
 6.1.2 Operating Temp. Range: -10 to 176°F (-25 to 80°C) 
 
 6.1.3 Battery life: Up to 1 year 
 
 6.1.4 Battery Type: CR2450 
 
 6.1.5 Download Cable: USB mini 
 
 6.1.6 Minimum Software Ver.: 10 
 
 6.1.7 Data Capacity: 32,000 (16,000 per channel) 
 
 6.1.8 RH Accuracy: +/-2%- 0 to 60%; +/-3%- 60 to 95% 
 
 6.1.9 Temperature Accuracy: +/- 0.8°F- 20 to 120°F, +/-1.8°F Remaining
 
 +/-0.44°C -7 to 49°C; +/-1.0°C Remaining 
 
 6.2 §P125- Temperature Data Logger, Digital 
 
 6.2.1 Operating RH conditions: 0-95% RH non-condensing 
 
 6.2.2 Operating Temp. Range: -10 to 176°F (-25 to 80°C) 
 
 6.2.3 Battery life: Up to 1 year 
 
 6.2.4 Battery Type: CR2450 
 
 6.2.5 Download Cable: USB mini 
 
 6.2.6 Minimum Software Ver.: 10 
 
 6.2.7 Data Capacity: 32,000 
 
 6.2.8 Temperature Accuracy: +/- 0.8°F- 20 to 120°F, +/-1.8°F Remaining
 
 /-0.44°C -7 to 49°C; +/-1.0°C Remaining 
 
 
 

[SOP 

 Standard Operating Procedure SOPNo | Rev 
 Verification of Digital Temperature and Humidity G-206 3 | Page 3 of4 
 Recording Devices 
 
 6.3 Other Data loggers with similar or better capabilities can be purchased and used. These
 
 types of devices will be added to this SOP as they become active. 
 
 7.0 Procedure 
 
 7A Every year before any verification procedure is performed a fresh battery will be
 installed into the data logger. If the data sampling is one minute or less the battery will
 
 need to be replaced at six month intervals. Experimental evidence shows that batteries
 have been lasting much shorter than 1 year. Batteries will be changed at least every two
 months to ensure they do not cause loss of data. 
 
 7.2 A temperature and humidity device calibrated against a NIST traceable standard will be
 used as a verification standard. 
 
 73 The calibration and certification of the temperature/humidity reference standard must be
 a performed by an outside laboratory or by the instrument manufacturer.
 
 7.4 Calibration verification will be documented using Form G-201-F2 in accordance with
 SOP G-201. 
 
 7.5 A single point verification will be performed using a controlled environment such as an
 incubation chamber or refrigerator. The controlled environment should be set to 35°C
 
 +/~ 5°C with environmental humidity or at a temperature within the refrigeration range
 of 4 to 8°C. Calibration of humidity should not be performed at refrigeration
 temperatures unless the humidity is controlled in the test environment.
 
 7.6 Both the internal reference standard and the digital recording device(s) must be fully
 
 equilibrated to the environment before the verification is initiated. The data range used
 in the performance verification should be clearly marked on the output graph.
 
 77 When the internal reference standard stabilizes, document the temperature and humidity
 reading and the time the verification is started and at multiple time intervals if the
 verification is performed for a specific time period (ex. Every 10 minutes for a 1 hour
 verification interval). The chamber must remain sealed during the verification process.
 
 For large climate controlled chambers multiple standard references should be used for
 the purpose of verifying the uniformity of the space used for the calibration.
 
 7.8 When the data collection is complete compare the device readings to the internal
 standard readings at similar time points and compare results to the specifications for
 that device. 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Verification of Digital Temperature and Humidity G-206 
 Page 4 of 4 
 Recording Devices 
 7.9 The specification for the verification is +/-2°C and +/- 4% humidity. The operation
 
 range for the device is -23 to 80°C and 0 to 95% humidity. 
 
 7.10 If the device is within the specified range at all monitoring points the unit is verified
 and a calibration sticker can be affixed to the unit before returning to service. Generate
 a report and add to the calibration log. 
 
 7.11 If the device fails the unit must be removed from service and replaced. Generate a
 
 report and add to the calibration log. 
 
 7.12 The calibration is good for 12 months. 
 
 8.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 04/26/10 | New - - | - | - |
| 1 | 03/07/13 | Updated SOP format and SOP number. Updated SOP format. Removed G-206-F1. Added G201-F2 and SOP G-201. Required NIST certified internal standards. Changed to | 13-131 | B. Johns |
| 2 | 01/27/16 | single point calibration in a tightly controlled environment. Introduced specifications. Adjusted SOP title. Updated responsibilities. | 15-1146 | R. Casabianca |
| 3 | 04/08/19 | Scheduled review: Updated battery requirement | 19-0235 | J. Maignan |