G-209

Calibration and Operation of the Sotax MT-50 Hardness Tester FORMATTED

Section G — Equipment Revision 0 5 pages

Original Document

Scanned document (image-only PDF)

Extracted Text

Searchable text extracted from PDF

1.0 Purpose 
 
 The purpose of this procedure ~is to provide detailed guidelines for the calibration and operation
 
 of the Sotax MT-50 hardness tester using the 5 kg weight and gage blocks.
 
 2.0 Scope 
 
 This procedure applies to the Sotax MT-50 hardness tester. This procedure provides
 
 comprehensive guidelines for the proper testing of tablet hardness as per USP <1217> Tablet
 
 Breaking Force. 
 
 3.0 Responsibility 
 
 3.1 It is the responsibility of the Metrologist or designee to keep this procedure aligned with
 current practices, and to calibrate the Sotax MT-50 hardness tester.
 
 3.2 It is the responsibility of Production (Compression) personnel to follow this procedure.
 
 3.3. ‘It is the responsibility of Production Management to ensure that this procedure is being
 
 followed. 
 
 3.4 ‘It is the responsibility of QC Laboratory Management to keep this procedure aligned
 
 with current practices. 
 
 4.0 Definitions 
 
 4.1 QC - Quality Control 
 
 4.2 USP- United States Pharmacopeia 
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Calibration and Operation of the Sotax MT-50 G-209 0 | Page 2 of 6 
 
 Hardness Tester 
 
 5.0 References 
 
 5.1 G-201-F2, Form, Calibration Certificate 
 
 5.2 USP <1217> Tablet Breaking Force 
 
 6.0 Use Procedure 
 
 6.1 Location and General Setup 
 
 6.1.1 Set the Sotax MT-50 hardness tester on a level surface free of vibration.
 
 6.1.2 Ensure the ambient temperature is between 4°C and 35°C, with a relative
 
 humidity between 10% and 75%, and no condensation. 
 
 6.1.3 Protect the unit from direct sunlight, dust, smoke, and corrosive fumes.
 
 6.2 Preparation for Use 
 
 6.2.1 Ensure the hardness tester is connected to a power source on a level.
 
 6.2.2 Turn on the unit and allow it to initialize. Allow unit to perform its automatic
 
 prestart check. 
 
 6.2.3 Verify that the pre-check was performed without issue. 
 
 6.2.4 Inspect the testing area and jaws for cleanliness; clean if necessary.
 
 6.3 Hardness Testing 
 
 6.3.1 Use the menu buttons to navigate to test then select “Easy Test”.
 
 6.3.2 Enter the number of samples to be tested next to the “H” field. Then press the
 
 green arrow. 
 
 6.3.3 Place a tablet in the test area according to product specifications.
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Calibration and Operation of the Sotax MT-50 G-209 0 | Page 3 of 6 
 
 Hardness Tester 
 
 6.3.4 If no specifications are provided, follow these guidelines: 
 
 6.3.4.1 Tablets are placed face up. 
 
 6.3.4.2 Round tablets can be oriented any way. 
 
 6.3.4.3 Elongated tablets should have the longer axis parallel to the jaw.
 
 6.3.5 Press the start arrow to initiate the test. 
 
 6.3.6 The left jaw will move to break the tablet, then retract, allowing for the next
 
 tablet to be placed. 
 
 6.3.7. Clean the test area as needed with a brush to remove debris.
 
 6.3.8 Repeat the test for the specified number of tablets in the sample group.
 
 6.3.9 Data recorded can be found in the list and statistic buttons at the top of the
 display. 
 
 6.4 Cleaning After Use 
 
 6.4.1 The hardness tester is a non-product contact surface. 
 
 6.4.2 Use a brush to remove broken tablets and clean the catch pan.
 
 6.4.3. Clean the instrument carefully with a slightly dampened rag with all-purpose
 
 cleaner. 
 
 6.4.3.1 Ensure that no liquid enters the unit and wipe it off with a soft cloth.
 
 6.4.3.2 Remove calcification by a suitable cleaning solvent. 
 
 6.4.3.3 Handle Plexiglas cover carefully and clean it with suitable cleaning
 
 solvent (no alcohol or acetone). 
 
 6.4.4 While cleaning, inspect the unit for wear and tear. Report any findings to the
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Calibration and Operation of the Sotax MT-50 G-209 
 Page 4 of 6 
 Hardness Tester 
 
 Metrologist or QC Laboratory Management. 
 
 7.0 Calibration Procedure 
 
 7.1 Setup for Calibration 
 
 7.1.1 Place the MT-50 hardness tester securely on a flat surface. 
 
 7.1.2 Connect the appropriate power cord and ensure the tester is powered on.
 
 7.1.3. Navigate to Service, Calib. 
 
 7.1.4 Ensure appropriate calibration settings are selected via the Calib. Settings
 
 option. 
 
 7.1.5 Select Hardness then rotate unit onto its left side. 
 
 7.1.6 Attach the calibration kit as described: 
 
 7.1.6.1 Remove the fixed jaw by loosening the two screws, taking care not to
 
 drop the jaw. 
 
 7.1.6.2 Fix the guide block in place of the crushing jaw (ensure center hole is
 
 aligned). 
 
 7.1.6.3 Insert the calibration plate into the guide block, ensuring the shaft is
 fully inserted into the calibration cube. 
 
 7.2 Calibration 
 
 7.2.1 Press the “Run” button and place the 5 kg reference weight on the calibration
 
 plate. Ensure the weight is centered correctly. 
 
 7.2.2 Press the “Run” button again to begin the calibration process.
 
 7.2.3, Wait for the message “Please wait...” to disappear and confirm the next
 
 
 

[SOP 

 Standard Operating Procedure SOP No Rev 
 Calibration and Operation of the Sotax MT-50 G-209 Page 5 of 6 
 
 Hardness Tester 
 
 calibration date displayed on the screen. 
 
 7.2.4 Press “Ok” to complete the calibration. Remove the reference weight and
 
 calibration kit. 
 
 7.2.5 Reinsert the fixed jaw and tighten both screws. 
 
 7.2.6 Return the MT-50 to its normal horizontal position. 
 
 8.0 Maintenance 
 8.1 Follow the preventive maintenance schedule as outlined in the Sotax MT-50 manual.
 
 8.2 Ensure all maintenance activities are documented and any issues are reported to
 
 management. 
 
 9.0 Revision History 

| Rev | Date | Description of Changes | CCR # | By |
|-----|----------|------------------------|-------|----|
| 0 | 06/11/24 | New procedure. N/A J. Nicholson | - | - |